Prepare to Scan:

Press Run/Stop to stop data collection. This is indicated by a red

1.

“Stopped” indication on the display.

 

2. Clear the memory buffer by pressing Math > Statistics > Clear Readings
(The number of stored samples is shown next to the Samples label)

Содержание SDM3000 Series

Страница 1: ...ing Excel Setup Configure the meter for the measurement of interest In this example we want to measure DCV 1 2 20 V fixed range 3 Set the measurement aperture to 1 PLC This is sometimes referred to as...

Страница 2: ...Stop to stop data collection This is indicated by a red 1 Stopped indication on the display 2 Clear the memory buffer by pressing Math Statistics Clear Readings The number of stored samples is shown...

Страница 3: ...the sample trigger to the desired number In this example we are going to set the sample count to 100 using the Up Down Left Right keypad NOTE Left adds decimal places right subtracts Up adds digits a...

Страница 4: ...Single button to start the acquisition When the meter is collecting data there will be a flashing green dot and the word trig on the display The sample count will also increment as each measurement i...

Страница 5: ...umber should equal to the setpoint in step 3 6 You can insert a USB stick FAT32 format into the front panel USB port and wait for the device to be recognized now or at any point from now till the actu...

Страница 6: ...Set current Path to External by pressing Browse and use the Up Down range arrows to highlight external and press select External will only appear when a USB memory stick FAT32 format is inserted into...

Страница 7: ......

Страница 8: ...ight arrows to select values 12 Remove the USB stick and open in a spreadsheet program Here you can see all of the voltages stored on the memory stick NOTE The data does not currently store timestamp...

Страница 9: ...total test time by the number of points collected and then manually enter the calculated timestamps for each point...

Отзывы: