SDG2000X_ServiceManual_SM
0202X-E01B.doc
28
To Set the DC offset
The following steps describe how to change offset to 1 Vdc.
1. Press the Offset/LowLevel softkey
The displayed offset is either the power-on value or the offset previously
selected. When changing the function, the same offset is used if the current
value is valid for the new waveform. To set the LowLevel for the waveform,
press the softkey again to switch to the LowLevel parameter.
2. Input the desired offset
Use the digital keypad to input the value directly, enter the value “
1
”.
Then press the corresponding softkey to select the desired unit. For example,
press
Vdc.