SIGLENT
SDS1000CFL Service Manual 21
Verify Test Results
To verify whether a test passes, whether the readings are within the limits, it is
necessary for you to record the readings in the Performance Test on Test
Record.
Self Test
This internal procedure is automatically performed every time the oscilloscope
is powered on. No test equipments are required. Verify that no error messages
are displayed before continuing with the procedure.
Self Calibration
You must perform the Self Calibration operation described in chapter 2. If the
environmental temperature changes by more than 5
°
, you must perform the
Self Calibration operation again.