SAP1000/SAP2500 Active Probe User Manual
7
including probe model, serial number, bandwidth, impedance, capacitance,
and attenuation ratio.
4. Select the channel to which the probe is connected. Set the oscilloscope
scale factor to 2 V/div. Set the oscilloscope offset factor to 0 V.
5. Measure the average voltage of channel 1, and the reading range should
be within ± (1.5%* full-screen r10mV). If the reading is beyond the
range, the check will not pass.
6. Change the scale factor of channel 1 to 1 V/div, 500 mV/div, 200 mV/div,
100 mV/div, 50 mV/div, 20 mV/div, 10 mV/div, and repeat Step 5 to check
the average voltage reading at each scale.
2.3 Quality Assurance
The oscilloscope has a 3-year warranty (1-year warranty for probe and
accessories) from the date of shipment, during normal use and operation.
SIGLENT can repair or replace any product that is returned to the authorized
service center during the warranty period. We must first examine the product to