Copyright ©Sigicom AB 2003-2015        Art.no ML089B02550-0En 
 

    
 

INFRA X20DM2 

Dust Monitor 
 

 

    
 
Art no. 080-02550-0 
 

Manual 

ver. D   

                  Valid for firmware version 1.1.0 

 

 
 

 

Содержание INFRA X20DM2

Страница 1: ...Manual Copyright Sigicom AB 2003 2015 Art no ML089B02550 0En INFRA X20DM2 Dust Monitor Art no 080 02550 0 Manual ver D Valid for firmware version 1 1 0...

Страница 2: ......

Страница 3: ...product This equipment falls under the Waste Electrical and Electronic Equipment Directive WEEE directive 2002 96 EC with its last update 2009 category 9 monitoring and control instruments Scrapped e...

Страница 4: ...8 3 2 Mounting 10 3 3 Start up instruction 10 3 4 Error checking 11 3 5 Failure recovery 11 4 Calibration and Maintenance 12 4 1 Calibration 12 4 2 Minor maintenance 12 4 3 Repair 15 5 Accessories 16...

Страница 5: ...A Master manual 1 1 About this manual This manual covers the following topics Chapter 1 Introduction Some introductory information about unpacking and safety and also an instruction how to assembly th...

Страница 6: ...tage immediately to Sigicom Record the instrument s serial number You will be asked to give this number in any X20DM2 related communication you may have with Sigicom 1 3 Safety Do never open the Osiri...

Страница 7: ...1 Take out the bushing see the picture above 2 Remove one of the white plastic cylinders and cut a slit in the bushing on its side towards the hole from the cylinder 3 Put the low voltage side of the...

Страница 8: ...6 Copyright Sigicom AB 2003 2015 4 Put back the bushing see the picture above 5 Connect the plugs inside the cabinet see the picture below...

Страница 9: ...Datalogger display 2 1 INFRA Field Monitoring system The INFRA contact placed on the bottom of the cabinet shall be connected to the INFRA Field Monitoring System by any of the available INFRA Sensor...

Страница 10: ...se the battery of Osiris is too weak to power the heating elements If external power is lost in this standard the interval data will be tagged with Overload S02 The inlet air is without heating The in...

Страница 11: ...ted interval file to the Data server Simultaneously an SMS or Server message alarm is possibly sent dependent on your settings The Upper threshold level can be set to any value in the interval 5 5000...

Страница 12: ...you want to run without external power you need to manually start the Osiris by pressing its ON Reset key The instrument can be run without external power for at least eight hours when Osiris is fully...

Страница 13: ...en using this standard the interval data will be tagged with Overload Also standard 2 need external power if we want to run for more than one day 3 5 Failure recovery During start up it might happen t...

Страница 14: ...and filter check 1 Stop sampling by doing Reg off at your INFRA Datalogger 2 Shutdown the Datalogger you must wait until all files have been uploaded 3 Remove external power 4 Remove the two connecto...

Страница 15: ...R iii press START three times password iv press ENTER to confirm filter change If not i press RESET twice 19 Connect your flow meter to the Osiris inlet located above the display 20 Hold the flow mete...

Страница 16: ...O ring 2 Replace filter Remember to reinsert the O ring Check the O ring is in good condition 3 Check the smaller O ring for the off center exhaust tube on the base 4 Replace the filter holder Make su...

Страница 17: ...Repair If sent to repair to Sigicom please carefully write down the description of the problem Here it is important to keep the cabinet as it is and send it in as one unit It is also recommended to co...

Страница 18: ...se for Dust monitor 2 art no 080 01887 0 Sensor drop cable 0 3m 1 0ft art no 080 01100 0 Sensor drop cable 1 0m 3 3ft art no 080 01101 0 Sensor drop cable 2 0m 6 6ft art no 080 01102 0 Sensor drop cab...

Страница 19: ...Copyright Sigicom AB 2003 2015 17 6 Contact and support Address Sigicom AB Alfred Nobels All 214 SE 146 48 Tullinge Phone 46 8 44 99 770 Internet www sigicom com E mail support sigicom se...

Страница 20: ...a in this section are given by Turnkey Instruments www turnkey instruments com Measurement range PM10 and TSP Channel 1 and 4 0 1 to 6000 g m3 PM2 5 and PM1 Channel 2 and 3 0 01 to 600 g m3 Detection...

Страница 21: ...Copyright Sigicom AB 2003 2015 19 A 4 Air flow Flow rate 600 cm3 min...

Страница 22: ...tor 4 Flow rate error in the dust monitor 8 Memory of the dust monitor is full 16 Photometer error of the dust monitor 32 Low battery error in the dust monitor 64 Over temperature inside the dust moni...

Страница 23: ...Copyright Sigicom AB 2003 2015 21 Appendix C Example of a Dust concentration monitoring The figure below shows monitored dust concentrations taken just outside Sigicom s office in Tullinge Sweden...