7.

Continuity

 

Testing

 

and

 

Resistance

 

Measurement

Attention !
Be absolutely certain that the device under test is 
voltage-free.

7.1

Continuity Testing

*

Verify that device under test is electrically dead.

*

Set the function selector switch to “     ” position.

*

If required null lead resistance by doing Zero adjustment below.

*

Connect the device under test as shown in diagram.

*

The measured resistance value will be displayed on the display.

*

The  instrument  generates  a  continuous  sound  signal  if 
the measured resistance value is less than the pre-setted 
“GoNoGO” value. Default value is 5

Ω

Refer to the “11. Menu structure” section for how to set the value
 for the limit value.

7.2

Low Ohm Measurement (

Ω

)

*

Verify that device under test is electrically dead.

*

Set the function selector switch to “ 

Ω

 ” position.

*

If required null lead resistance by doing zero adjustment below.

*

Connect the device under test as shown in diagram

*

The measured resistance value will be displayed on the display.

*

The instrument generates a continuous sound signal if the
measured resistance value is less than the pre-setted 
“GoNoGO”  value. Default value is 0

Ω

Refer to the “11. Menu structure” section for how to set the 
value for the limit value.

7.3

Zero adjustment

When  measuring  very  low  resistance  values,  you  can  eliminate 
resistance of leads and contact resistance by zero adjustment.

*

Connect the test leads to instrument and connect the free ends of
 the leads to one another.

*

Wait for reading to stabilize.

*

Briefly  press  the  “TEST”  push  button.  The  instrument
acknowledges  zero  adjustment  by  a  sound  signal  and 
the LCD shows “0.00” (+1digit) and the symbol “ZERO” (17) is
displayed. The resistance measured at the instant the push button
is pressed is used as reference value (0-9.99

Ω

). It is automatically

deducted from the values measured thereafter.

*

Zero adjustment can be cleared.

Ø

 By pressing the “TEST” button once again.

Ø

 By switching the instrument o

ff

.

Also zero adjustment is cleared when Auto shutOFF operates.

Additional features available in Ohm and Continuity ranges.

1.

Pre-selectable limit checks (Go/Nogo option)

2.

Storing the measured resistance value.

3.

Minimum and Maximum value “MIN/MAX” storage.

11

10

Continuity

  

Test

-

0 0

0

M

W

50V

100V

250V

500V

1KV

V

+

-

0L

M

W

50V

100V

250V

500V

1KV

V

+

Содержание Insu 10

Страница 1: ...am Tinsley Instrumentation Inc 3105 Creekside Village Drive Suite No 801 Kennesaw GA 30144 USA Contact No 1 404 736 4903 E mail Id psk sifamtinsley com Web www sifamtinsley com User Manual Analog Digital Insulation Tester Insu 10 ...

Страница 2: ...poses the operator the system and the measuring instrument to danger The use of repaired fuses or short circui ng of the fuse holder is prohibited 13 3 Housing No special maintenance is required for the housing Excessive contamination has an adverse effect on isolation and reduces input resistance The surface must be kept clean for this reason Use a slightly dampened cloth for cleaning Oil free pe...

Страница 3: ...SE appears on display If display OL Fuse is OK If display FUSE Fuse is blown and needs replacement Note Once FUSE is displayed all the functions are inhibited Immediately switch the Instrument OFF and replace the fuse 18 3 1 Liquid Crystal Display 10 Symbols for displaying the 2 On Off push button selected functions 3 Push button 11 Display for selected function 4 Push button 12 Display for the un...

Страница 4: ...STOP WATCH mode or in MENU structure During an Insulation test connections must not be touched After insulation tests capacitive circuits must be allowed to discharge before disconnecting test leads Test leads and crocodile clips must be in good condition clean and with no broken or cracked insulation u 4 17 Power Supply Battery 6 x 1 5V cells IEC LR6 Nickel cadmium rechargeable cells may be used ...

Страница 5: ...the measuring circuit and ensure high voltage is not available on leads Automatic turn off Your instrument turn off automatically if any of the keys or the function selector switch have not been activated for about 10 minutes in insulation resistance range and 5 minutes in other ranges How to prevent automatic TURN OFF In order to prevent automatic TURN OFF select CONTINUOUSLY ON mode For this pus...

Страница 6: ...ication and Digital display and with display of unit of measured quantity and functions Analog Display LCD scale with bar graph pointer Scale Length 47 mm Scaling 0 30 with 30 graduations Over range Indication By triangle 13 Digital Display Char Height 7 segment digits 12mm Number of Digits 3 digit for MΩ and Ω 4 digit for Stop watch Over range Indication OL ON 2 0 R Insulation Resistance 1 MΩ U 5...

Страница 7: ...ow the last reading for few seconds even after releasing the TEST key Note The instrument s batteries are depleted during insulation resistance measurement Only press and hold the TEST key as long as it is necessary to take the reading in the manual measurement mode Continuous measurement as described below should be performed if absolutely necessary 7 14 11 1 Description of Menu Items 11 1 1 Sett...

Страница 8: ... is set it is applicable to all the 5 ranges of Insulation measurement n Select the required value of the time for which the test is to be carried out n Briefly Press the TEST key The insulation resistance test carries on for the time interval selected n The test terminates automatically after the selected time has elapsed n The display then flashes the value of measurement at the end until the Key ...

Страница 9: ...u can still read the actual measured value Apply the measured quantity to instrument prior to activating MIN MAX function The MIN MAX function is switched off when 1 Key is pressed for approximately for 1 Sec OR 2 Then function selector switch is operated OR 3 When the instrument is turned OFF and ON again 9 12 8 Voltage Measurement V AC DC You can use this instrument for voltage measurement from ...

Страница 10: ...nal if the measured resistance value is less than the pre setted GoNoGO value Default value is 0Ω Refer to the 11 Menu structure section for how to set the value for the limit value 7 3 Zero adjustment When measuring very low resistance values you can eliminate resistance of leads and contact resistance by zero adjustment Connect the test leads to instrument and connect the free ends of the leads ...

Отзывы: