4.1.15 

 

                                         SmartTrak® 50 Series Low & Medium Flow Quick Install Guide 

 

 

47-0443 C

 

 

3.  Establish Communications 

Close the window using the X button in the upper right corner of the dialog box. The communications are set up, and the 
unit should be communicating with the computer, as evident by changing numbers to the right of the Comm Port and a 
Green status light.  If the numbers are not changing, click ‘Get Parameters.’ 

 

 

 

RESET NOTE: 

If you are not able to establish communication with your instrument, be sure you have the proper ComPort 

selected to match your PC port then hit the “Reset Host Port to Default” button in the Port Dialog Box and power down 
your instrument.  Next, power up the instrument again and hold the zero pushbutton (located inside your instruments side 
plate behind the silver socket head cap screw) for a minimum of one second after powering up your instrument. This will 
reset both the host software and your instrument to the default settings. Click on the ‘Get Parameters’ button and 
communication should be established. 
 

4. Read and Perform Adjustments 

With communications established you can now read flow, gas, units, setpoint and full scale, along with communications and 
valve status.  You can also adjust setpoint, zero and span.  Please consult the manual for a more detailed explanation. 

 

 

LIMITED WARRANTY POLICY- REGISTER ONLINE  

All Sierra products are warranted to be free from defects in material and workmanship and will be repaired or 
replaced at no charge to Buyer, provided return or rejection of product is made within a reasonable period but 
no longer than one (1) year for calibration and non-calibration defects, from date of delivery. To assure warranty 
service, customers must register their products online on Sierra’s website. Online registration of all of your 
Sierra products is required for our warranty process. Register now at 

www.sierrainstruments.com/register

Learn more about Sierra’s warranty policy at 

www.sierrainstruments.com/warranty

 

 

QUESTIONS?  We are here to help when you need us! 

Email Tech Support:

 

[email protected]

 

24 Hour Website Service:

  

www.sierrainstruments.com

 (Click “Sales & Service” button). 

Tech Support:  
N. America:

 800.866.0200 or 831.373.0200

  

Europe:

+ 31 72 5071 400

  

Asia Pacific:

 + 8620 3435 4870 

Содержание SmartTrak 50 Series

Страница 1: ...m of your instrument If the gas contains any moisture use an appropriate dryer or desiccant Particles larger than 10 micron and moisture may damage your instrument 3 Mount according to Data Label orientation Horizontal flow is preferable and factory default unless the factory calibration was performed specifically for vertical flow upward or downward as listed Orientation is listed on Data Label a...

Страница 2: ...ge the span change setpoint or power up setpoint or change the address of the unit software is talking to RS 485 only Locate this disk At this point EXIT OUT OF ANY OPEN APPLICATIONS BEING RUN ON YOUR COMPUTER 1 Insert the SmartTrak 50 Series Software CD into your CD ROM 2 Open My Computer on your desktop 3 Open the CD Named SmartTrak 50 Series on your D drive 4 Run setup exe 5 Follow the instruct...

Страница 3: ...on o Expand Root Menu Ports COM LPT by Left Clicking arrow left of name It will look like this o Identify port being used by the 50 series Windows 7 o Right Click on Computer in the Start Menu then Left Click Properties from there Device Manager is listed in the left hand frame menu NOTE Device Manager links look like this in Win7 o Left Click Device Manager dialog button and expand the root menu ...

Страница 4: ...ablished 4 Read and Perform Adjustments With communications established you can now read flow gas units setpoint and full scale along with communications and valve status You can also adjust setpoint zero and span Please consult the manual for a more detailed explanation LIMITED WARRANTY POLICY REGISTER ONLINE All Sierra products are warranted to be free from defects in material and workmanship an...

Отзывы: