                                         SmartTrak® 50 Series Low & Medium Flow Quick Install Guide 

 

 

47-0443 B

 

 

SmartTrak® 50 Series Low & Medium Flow 

Quick Install Guide 

 

INTRODUCTION

 

Congratulations on your purchase of the SmartTrak® 50 Series digital gas mass flow meter/controller designed to meet the 
low-cost budgetary requirements of OEM’s while maintaining excellent quality and performance.  This reference guide is 
not meant to replace the Instruction Manual included on the CD-ROM in your shipment, but rather is a great supplement to 
it that will help you install and operate your instrument and begin to measure or control mass flow. 
 

SAFETY

 1

st

 

1.

 

Apply power only after reviewing the wiring diagrams on page 2 of this guide or in the Instruction Manual. 

2.

 

Apply gas flow only after checking plumbing connections for leaks. 

3.

 

NEVER TEST FOR LEAKS WITH LIQUID LEAK DETECTOR.

 If liquid seeps into the electronics or the hidden sensor 

compartment, the instrument may be damaged. Instead use a pressure-decay test (if liquid MUST be used at all, limit it 
to the fittings and keep it off the body of the instrument).  

 

INSTALLATION

 

1.

 

Consult the instrument’s Data Label (on the back of the instrument)

 for ALL proper operating parameters. If the 

information on the Data Label does not match your process conditions, contact your Representative or Sierra 
Instruments Customer Service at the contact information on the footer of the sheet. 

2.

 

Install a 10 micron in-line filter upstream of your instrument.

 If the gas contains any moisture, use an appropriate 

dryer or desiccant. Particles larger than 10 micron and moisture may damage your instrument. 

3.

 

Mount according to Data Label orientation.

 Horizontal flow is preferable and factory default unless the factory 

calibration was performed specifically for vertical flow upward or downward, as listed. Orientation is listed on Data 
Label and on the Calibration Certificate. 

4.

 

This instrument is not a loop-powered device!  Do NOT apply power to the 4-20 mA / 0-5 VDC output or setpoint 
connections. 

5.

 

Wire your instrument per the diagramed pin-out

 on the back of this card and in Chapter 2 of the Instruction Manual. 

For convenience of wiring, there are two DB9 connector ports; the unused DB9 is capped off.  Apply power to your 
choice of port and cap unused port.  It is recommended you allow 15 minutes for the instrument to warm up for 
maximum accuracy. 

 

OPERATION

 

1.

 

Power the unit.

 If you are using the Sierra provided power supply, it is recommended you attach the D-sub connector 

to the 50 Series before plugging the adapter into the wall. The unit may take longer to start than normal if otherwise. 
This will not damage the unit in any way.  

2.

 

Our Mass Flow Controllers with a digital setpoint option 

are shipped with a 

zero setpoint

 in the power-up setpoint box 

in the software. 

For safety considerations, it is recommended

 YOU 

confirm this prior to applying gas to the unit.

 

3.

 

Apply the gas

 listed on the Data Label

 to the inlet at the recommended pressure (listed on the Data Label/Calibration 

Certificate).  Your 50 Series will begin to measure mass flow. This will be displayed on the user software or local display. 

4.

 

You will need a setpoint input in order to control flow. 

Do so carefully!

 You can digitally input a setpoint using the user 

software or control directly with your analog setpoint source if you ordered the analog option. 

NOTE:

 The unit ships in 

Содержание SmartTrak 50 Series

Страница 1: ...m of your instrument If the gas contains any moisture use an appropriate dryer or desiccant Particles larger than 10 micron and moisture may damage your instrument 3 Mount according to Data Label orientation Horizontal flow is preferable and factory default unless the factory calibration was performed specifically for vertical flow upward or downward as listed Orientation is listed on Data Label a...

Страница 2: ...ge the span change setpoint or power up setpoint or change the address of the unit software is talking to RS 485 only Locate this disk At this point EXIT OUT OF ANY OPEN APPLICATIONS BEING RUN ON YOUR COMPUTER 1 Insert the SmartTrak 50 Series Software CD into your CD ROM 2 Open My Computer on your desktop 3 Open the CD Named SmartTrak 50 Series on your D drive 4 Run setup exe 5 Follow the instruct...

Страница 3: ...on o Expand Root Menu Ports COM LPT by Left Clicking arrow left of name It will look like this o Identify port being used by the 50 series Windows 7 o Right Click on Computer in the Start Menu then Left Click Properties from there Device Manager is listed in the left hand frame menu NOTE Device Manager links look like this in Win7 o Left Click Device Manager dialog button and expand the root menu ...

Страница 4: ...ablished 4 Read and Perform Adjustments With communications established you can now read flow gas units setpoint and full scale along with communications and valve status You can also adjust setpoint zero and span Please consult the manual for a more detailed explanation LIMITED WARRANTY POLICY REGISTER ONLINE All Sierra products are warranted to be free from defects in material and workmanship an...

Отзывы: