Chapter 10: Annual Maintenance and Calibration 

 

Your CalTrak SL 800 is engineered to provide years of reliable service, with appropriate care and maintenance. Sierra 
recommends an annual calibration, to help ensure the best possible flow measurements and to provide an audit trail for 
those applications subject to regulatory requirements. If you should encounter any problems with your CalTrak SL 800, 
immediately contact Customer Service and provide a detailed description of your situation, including CalTrak SL 800 
model and serial number, information about the flow source and the current calibration setup, environmental conditions 
during the test, the flow point or points that you’re checking and an explanation of the issue you’re experiencing. 

Recertification 

Your CalTrak SL 800 primary piston prover is a precision measuring standard comprised of moving parts that are ma-
chined to extremely close tolerances. Additionally, various environmental factors, product wear, drift of the temperature 

sensors and pressure transducers, or inadvertent damage may adversely affect your measurement accuracy or general 
performance.

 

For  these  reasons,  Sierra  highly  recommends  having  your  CalTrak  SL  800  annually  verified  in  order  to  ensure  its 

measurement integrity. For those applications subject to regulatory or ISO requirements, verification by oan accredited 
laboratory provides you with a defensible audit trail of the highest quality.

 

 

Returning Equipment to Factory 

Please contact Sierra for a service quote before sending your CalTrak SL 800 to our factory.  

 
If you are sending in your CalTrak for repair or evaluation (rather than elective re-certification), contact Sierra for technical 

support or troubleshooting assistance prior to shipping the unit.  Provide us a detailed description of your issues. If we are 
unable to resolve the situation by phone or email, we will issue you an RMA (return merchandise authorization) number.  
Follow the instructions for returning your instrument for service as noted below.  

 

 

 

Shipping 

When shipping your CalTrak, be sure to follow some simple guidelines to avoid costly damage to your property.   

 

Pack your instrument carefully.  Use adequate packing material. Whenever possible, use the original 
packing and foam or bubble wrap that came with your CalTrak (packing peanuts NOT recommended). 

Or use a Sierra CalTrak Pelican carrying case, which provides a hard case shell for protection of your 

valuable equipment. If you do not already have a Pelican case, visit us at www.sierrainstruments.com  

for more information on obtaining one. 

 

Include a copy of the RMA form (complete with Sierra supplied RMA number) with the unit(s).    

 

 

Use a major freight carrier (e.g., FedEx, UPS) that supplies tracking numbers. 

 

Insure your CalTrak. Sierra is not responsible for damage occurred during transit. 

 

Understand our mutual shipping obligations. Sierra is responsible for shipping cost only if the issue is 

product related and the CalTrak is still under warranty. 

 

RMA Note – Returning Unit for Service 

Sierra will not evaluate or service your instrument without an RMA number.   Go to 

http://www.sierrainstruments.com/rma

 to complete an RMA. 

 

Содержание CalTrak 800 Series

Страница 1: ...1 CalTrak 800 High Accuracy Modular Primary Standard Gas Volume Flow Calibrator Cal Labs Industry Applications Instruction Manual Models SL800 Part Number IM Cal SL800 Rev F 4 16...

Страница 2: ...879 8521 Fax 8621 5879 8586 COPYRIGHT SIERRA INSTRUMENTS 2016 No part of this publication may be copied or distributed transmitted transcribed stored in a retrieval system or translated into any human...

Страница 3: ...rdance with published CE directives Warning Do not power the flow meter with the sensor remote if applicable wires disconnected This could cause over heating of the sensors and or damage to the electr...

Страница 4: ...e consistent with factory recommendations Installation information can be found in Chapter 2 of this manual and troubleshooting information can be found in Chapter 10 If the problem persists after fol...

Страница 5: ...oving Flow Cells 14 Chapter 5 Factory Default Settings 15 Chapter 6 Connecting the CalTrak 800 to Gas Flow Source 16 Chapter 7 Taking Readings 17 Chapter 8 Using Sensor Factors for MFC and MFM Calibra...

Страница 6: ...in an averaging sequence The versatile CalTrak SL 800 can be used to measure the gas flow rates at atmospheric pressure 7mmHg Its Proven CalTrak Technology features a near frictionless piston and flo...

Страница 7: ...32 Serial Cable USB Cable Certificate of Calibration CalTrak SL 800 Flow Cell Dust Caps Certificate of Calibration External Inlet Filter SL 800 3 only Carrying cases and accessories are available for...

Страница 8: ...signed to operate using only AC power this unit does not have a battery Ensure that your unit is plugged in prior to turning on Communications Ports The CalTrak SL 800 includes a USB B port and RS 232...

Страница 9: ...creen and all submenus You can return to this screen at any time during operation by pressing the Home button in the lower left corner of the screen Measure Screen The Measure screen allows you to cap...

Страница 10: ...ensor factor PTVM standardized or volumetric flow standardization temperature and reading average number When you are done changing thesevalues pressing CONFIRM will take you to the Measure screen The...

Страница 11: ...rial number firmware revision date of last cell calibration cycles run since lastcalibration calibration number model of installed cell and cell serial number Provide these numbers when contacting cus...

Страница 12: ...he measurements On our highest accuracy instruments we allow for the adjustment of the Piston Tare Value with the Piston Tare ValueMultiplier PTVM When using the instrument with gas species other than...

Страница 13: ...he standardization temperature The standardization pressure is set at 760mmHg 14 7 psi 1 atm 1013 25 mbar Change the standardization temperature by selecting this option and changing the value using t...

Страница 14: ...celldue to its larger diameter Replace this adapter ring when using the other cells Installing Flow Cells 1 Ensure blue adapter ring is installed for using all cells except the 75 Remove the ring if i...

Страница 15: ...s To return to factory default settings at any time press Reset followed by Save from the Main Menu Parameters Factory Settings Optional Settings No of Readings in an Averaging Sequence 10 1 100 Atmos...

Страница 16: ...flow source connect to the outlet fitting The CalTrak SL 800 is designed to be used at ambient pressures only This is easily accomplished by leaving the outlet of the flow cell open to atmosphere for...

Страница 17: ...ycle through options for single burst or continuous readings Press Start button to begin measurements Press the Stop button to stop current flow reading and open valve Setting User Preferences The Cal...

Страница 18: ...MFM manufacturer can provide the appropriate sensor factor value The user can also apply this flow correction factor to the CalTrak SL 800 readings To enable your CalTrak SL 800 to scale its actual f...

Страница 19: ...button to reset and install the cell Piston Error This message means that the piston did not return to the bottom of the cell properly during operation Hit the home button to reset If you continue to...

Страница 20: ...Please contact Sierra for a service quote before sending your CalTrak SL 800 to our factory If you are sending in your CalTrak for repair or evaluation rather than elective re certification contact Si...

Страница 21: ...21 Ship the unit s to the following address Sierra Instruments Inc Attention Factory Service Center 5 Harris Court Building L Monterey CA 93940 USA RE RMA your number...

Страница 22: ...22 Chapter 11 Storage Protecting Your CalTrak When Not In Use If you need to store your CalTrak SL800 for an extended period please follow these guidelines Always store it in a clean dry place...

Страница 23: ...n 70 humidity Usage Flow Mode Suction and Pressure Configuration Base with modular interchangeable flow cells five Temperature Pressure Sensors Yes in the flow stream of each interchangeable flow cell...

Страница 24: ...uct is made within a reasonable period but no longer than one 1 year for calibration and non calibration defects from date of delivery To assure warranty service customers must register their products...

Отзывы: