Sierra Accu-Mass 730 Скачать руководство пользователя страница 11

Series 730 Instruction Manual

IM-73-E

9

Problem

Possible Cause

Solution

Flow measurement is
erratic or fluctuating

Very erratic or non-uniform flow

Follow installation requirements
shown in Figure 2

Flow meter installed with less than re-
quired minimum pipe diameters up-
stream and downstream of the sensor

Follow installation requirements
shown in Figure 2

Probe is not secured to a solid base

Secure probe to a non-vibrating solid
mount

Sensor component broken

Return to factory for replacement

Malfunction in system
electronics

Return to factory for evaluation

Moisture present in gas flow

Install a water trap or filter upstream
of the flow sensor

Velocity measurement
seems too high or low

Sensor assembly not aligned correctly
to flow

Correct alignment with the flow indica-
tor pointing downstream in the direc-
tion of flow

No response to flow from
sensor assembly

No power

Turn on power to the flow meter

Flow profile distortions

Try to find another location for the
meter

Extremely turbulent flow

Do not place the meter near a ventila-
tor, static mixer or valve

Sensor failure

Return to factory for repair

Malfunction in system
electronics

Return to factory for evaluation

System will not zero

Out of calibration

Return to factory for calibration

Probe tip is broken

Return to factory for repair

4-20 mA output circuit not
indicating 4 mA at zero
flow

Improper current loop resistance; total
current loop resistance must be no
less than 150 Ohms and no greater
than 450 Ohms

Install load resistor or use larger
gauge wire

Output pegs plus or minus

Flow body not plugged in

Plug in both connectors

Malfunction in system
electronics

Return to factory for evaluation

Содержание Accu-Mass 730

Страница 1: ...Part Number IM 73 AM Revision E 05 04 5 Harris Court Building L Monterey CA 93940 831 373 0200 800 866 0200 Fax 831 373 4402 http www sierrainstruments com Sierra Instruments b v Bolstoen 30A 1046 AV...

Страница 2: ...publication may be copied or distributed transmitted transcribed stored in a retrieval system or translated into any human or computer language in any form or by any means electronic mechanical manua...

Страница 3: ...bstructed Flow Requirements 5 Installation 6 Wiring Connections NEMA 2 Enclosures 7 Wiring Connections NEMA 4 Enclosures 7 Output Signal Wiring 8 Troubleshooting the Flow Meter 8 Returning Equipment t...

Страница 4: ...meter repair verify that the line is de pressurized Caution Always remove main power before disassembling any part of the flow meter Caution Printed circuit boards are sensitive to electrostatic disc...

Страница 5: ...of a 304 stainless steel probe which is inserted in the gas stream The sensor consists of two sensing ele ments a velocity sensor and a temperature sensor which automati cally corrects for changes in...

Страница 6: ...ial Do not return any equipment to the factory without first contacting Sierra Customer Service Technical Assistance If you encounter a problem with the flow meter review the configura tion informatio...

Страница 7: ...measured 5 For remote installations verify the supplied cable length is sufficient to connect the flow meter sensor to the remote electronics Do not extend or shorten the supplied cable be tween the...

Страница 8: ...umber of diameters of straight pipe required downstream of the flow meter Flow Flow Flow Flow Flow Flow Figure 2 Recommended Pipe Length Requirements for Installation Installation The following instru...

Страница 9: ...input power 300 mA load maximum to pin 1 and power ground to pin 5 See Output Sig nal Wiring for signal wiring instructions Figure 3 NEMA 2 CE Enclosure Wiring Connections Warning All wiring procedure...

Страница 10: ...s 3 Verify the flow meter is installed with the correct number of up stream and downstream pipe diameters as shown in Figure 2 4 Verify the flow direction indicator is correctly aligned pointing downs...

Страница 11: ...flow Correct alignment with the flow indica tor pointing downstream in the direc tion of flow No response to flow from sensor assembly No power Turn on power to the flow meter Flow profile distortion...

Страница 12: ...ts com Or contact Customer Service at service sierrainstruments com Email or 800 866 0200 or 831 373 0200 in the US or 31 0 20 6145810 in Europe Return shipments to USA Headquarters Sierra Instruments...

Страница 13: ...DC 300 mA Output Signal Linear 0 5 VDC or 0 10 VDC proportional to mass flow rate 1000 Ohms minimum load resistance or linear 4 20 mA proportional to mass flow rate 400 Ohms maximum resistance Perform...

Страница 14: ...ite Signal 3 Red DC input Power 4 N C 5 N C 6 Green Power Ground 7 Brown or Orange Bridge Black Connector Female Mating Connector Solder Side No Color Signal 1 Red DC input Power 2 White Signal 3 Blac...

Отзывы: