Error Correction
5 - 7
Siemens AG
SPL5-330.840.01
Page 7 of 12
Uroskop ACCESS
Medical Solutions
Rev. 01
03.03
CS SD 24
Endoscopy Option
Example of the test configuration with chrome-plated pliers, contrast OK.
Arrow 1:
The reflective areas of the pliers appear almost too bright and glaring and have a notice-
able white border, which indicates the beginning of camera overmodulation. The reflective
areas still indicate noise in the image.
Arrow 2:
When the monitor is set correctly, the structure of the pliers is visible without any satura-
tion effects.
Fig. 5