![Siemens SIWAREX WP251 Скачать руководство пользователя страница 97](http://html.mh-extra.com/html/siemens/siwarex-wp251/siwarex-wp251_manual_374538097.webp)
Scale parameters and functions
8.11 DR 10 Load cell parameters
SIWAREX WP251
Manual, 12/2015, A5E37203357A
97
Variable
Note
Type
Length
(bytes)
Rw Default Min. Max.
Modbus
registers
Application
Information about which
application the DR belongs
to
USHORT
2
r
105
-
-
1402
Version identifier
Current data record version
information
USHORT
2
r
1
1
65635
1403
Reserve
Reserve
USHORT 2
rw
1
-
-
1404
Reserve
Reserve
USHORT 2
rw
0
-
-
1405
Number of support points
USHORT 2
rw
0
0
8
1406
Reserve
Reserve
USHORT 2
rw
0
-
-
1407
1)
Characteristic value of the
connected load cell(s). The
mean value (mV) is used if
there is more than one cell.
FLOAT
4
rw
2
> 0.1 10
1408
Reserve
Reserve
FLOAT
4
rw
0
-
-
1410
(Page 98)
1)
Nominal load of one load cell FLOAT
4
rw
60
0
Specifi-
cation
from
DR 3
1412
Reserve
Reserve
FLOAT
4
rw
0
-
-
1414
Reserve
Reserve
FLOAT
4
rw
0
-
-
1416
Reserve 2
Reserve
SHORT
2
rw
0
-
-
1418
Reserve 3
Reserve
USHORT 2
rw
0
-
-
1419
Reserve 4
Reserve
FLOAT
4
rw
0
-
-
1420
Load cell manufacturer
string header
String header
UBYTE[2] 2
rw
24,24
1422
Load cell manufacturer Load cell manufacturer
CHAR[24] 24
rw
0.0
1423
Load cell order number
String header
String header
UBYTE[2] 2
rw
24,24
1435
Load cell order number Load cell order number
CHAR[24] 24
rw
1436
1)
Parameter for calculation of calibration points with theoretical calibration
8.11.2
Number of support points
If no anchor points are used, the number of support points is equal to the number of load
cells.
If anchor points are used in addition to load cells, the number of support points is equal to
the total number of load cells and fixed support points.
Example
A tank is mounted on three load cells → Number of mechanical support points = 3
Содержание SIWAREX WP251
Страница 28: ...Mounting 5 3 Mounting on the SIMATIC S7 1200 SIWAREX WP251 28 Manual 12 2015 A5E37203357A ...
Страница 40: ...Connection 6 12 Attachment of calibration protection plate SIWAREX WP251 40 Manual 12 2015 A5E37203357A ...
Страница 56: ...Commissioning 7 5 Service with the SIWATOOL program SIWAREX WP251 56 Manual 12 2015 A5E37203357A ...
Страница 162: ...Scale parameters and functions 8 38 DR 48 date and time 2 for Modbus SIWAREX WP251 162 Manual 12 2015 A5E37203357A ...
Страница 212: ...Technical specifications 13 3 Approvals SIWAREX WP251 212 Manual 12 2015 A5E37203357A ...
Страница 218: ...List of abbreviations B 1 List of abbreviations SIWAREX WP251 218 Manual 12 2015 A5E37203357A ...