Functions
9.16 Factory parameters
SITRANS TR200/TR300
Operating Instructions, 06/2010, A5E01071992-03
57
9.16
Factory parameters
Using the menu command Device ➜ Factory trim ➜ Reset factory settings, the configuration
of the transmitter is reset to its default values.
Note
Factory reset
The menu command Device ➜ Factory trim ➜ Reset factory settings always resets the
parameters of the transmitter to the default values in the table listed below.
This factory reset has the same effect for "devices supplied ex stock" or "devices set
customer-specifically".
Furthermore, resetting the transmitter to its default values also resets a customer-specific
digital-to-analog converter trim and sensor calibration (one point trim or two point trim).
Once a factory reset has been performed, the transmitter has the following configuration:
Parameter
Reset to value
DAY
Is not reset
Description
Is not reset
Message
Is not reset
Serial number
Is not reset
Installation date (electronics)
Is not reset
Sensor class
Resistance thermometer
Sensor type
Pt100 DIN IEC 751
Interface
Standard connection
Sensor connection
Three-wire input
Sensor factor
1.00
Sensor offset 1
0.00 °C
Start of scale value
0 °C
Full scale value
100 °C
Unit
°C
Break monitoring
ON
Short circuit monitoring
ON
Short circuit limit
10 Ω
Lower end point analog output
Is not reset
Upper end point analog output
Is not reset
Alarm value
Is not reset
Linearization type
Linear to temperature
Attenuation
0.00 s
Runtime meters PV
Are all reset to 0 h
Runtime meters, field device
Are not reset
Min/max pointers PV
Are all reset to 0
Min/max pointers for electronics temperature
Are not reset
Manufacturer data sensor
Are not reset
Содержание sitrans tr300
Страница 7: ...Table of contents SITRANS TR200 TR300 6 Operating Instructions 06 2010 A5E01071992 03 ...
Страница 13: ...General safety notes 2 5 Measures SITRANS TR200 TR300 12 Operating Instructions 06 2010 A5E01071992 03 ...
Страница 21: ...Installation SITRANS TR200 TR300 20 Operating Instructions 06 2010 A5E01071992 03 ...
Страница 43: ...Commissioning 8 2 LED operating indicator SITRANS TR200 TR300 42 Operating Instructions 06 2010 A5E01071992 03 ...
Страница 69: ...Service and maintenance SITRANS TR200 TR300 68 Operating Instructions 06 2010 A5E01071992 03 ...
Страница 81: ...Dimension drawings SITRANS TR200 TR300 80 Operating Instructions 06 2010 A5E01071992 03 ...
Страница 83: ...Spare parts and accessories SITRANS TR200 TR300 82 Operating Instructions 06 2010 A5E01071992 03 ...
Страница 87: ...List of abbreviations B 1 Abbreviations SITRANS TR200 TR300 86 Operating Instructions 06 2010 A5E01071992 03 ...
Страница 93: ...Glossary SITRANS TR200 TR300 92 Operating Instructions 06 2010 A5E01071992 03 ...