A5E32106153
SITRANS Probe LR – QUICK START MANUAL
Page EN-3
m
mmmm
Englis
h
Power
• Maximum 30 V DC
• 4 to 20 mA
Approvals
• General
CSA
US/C
, FM, CE, C-TICK
• Radio
Europe (R&TTE), FCC, Industry Canada
• Hazardous Intrinsically Safe
(Europe)
ATEX II 1 G Ex ia IIC T4 Ga
(US/Canada)
FM/CSA
1
: (barrier required)
Class I, Div. 1, Groups A, B, C, D
Class II, Div. 1, Groups E, F, G
Class III
T4
(International) IECEx SIR 13.0005X Ex ia IIC T4 Ga
(Brazil)
INMETRO: DNV 12.0068 X
Ex ia IIC T4 Ga
IP67/IP68
-40 ºC
≤
T a
≤
+80 ºC
DNV #OCP 0017
ABNT NBR IEC 60079-0:2008 e
ABNT NBR IEC 60079-11:2009 e
ABNT NBR IEC 60079-26:2008
Non-incendive
(US)
FM
2
:
Class I, Div. 2, Groups A, B, C, D
T5
• Marine
Lloyd’s Register of Shipping
ABS Type Approval
Mechanical
• Process
Connections: threaded connection
1.5” NPT, BSP, or G (BS EN ISO 228-1)
• Antenna:
polypropylene rod
hermetically sealed construction
Process
• Temperature (at process connection):
40 to +80 °C (
40 to +176 °F)
• Pressure (vessel):
maximum 3 bar, gauge (43.5 psi, gauge)
Nominal 24 V DC at max. 550 Ohm
1.
See
Wiring Drawing: Intrinsically Safe (FM )
on page 2 of Appendix A, for drawing number 23651611,
or
Wiring Drawing: Intrinsically Safe (CSA)
on page 3 of Appendix A, for drawing number 23651621.
2.
See
Wiring Drawing: Non-incendive (FM)
on page 1 of Appendix A, for drawing number 23650537.
Note:
Approved dust-tight and water-tight conduit seals are required for outdoor Type 4X /
NEMA 4X, Type 6 / NEMA 6, IP67, IP68 locations.
Note:
EN61000-4-3 (CE EMC) testing was conducted on the SITRANS Probe LR while
mounted in a metallic vessel.