
Page EN-2
SITRANS LC300 – QUICK START MANUAL
A5E32268590
mmmmm
English
Definitions
SITRANS LC300
SITRANS LC300 is a cost-effective instrument for level measurement in applications such as
the processing of food and beverages, pharmaceuticals, detergents and pet food. It performs
in liquids, bulk solids and slurries, including viscous (conductive or non-conductive) materials,
even in challenging environments involving vapour and dust.
LC300 is a 2-wire instrument combining a sophisticated, yet easy-to-adjust, microprocessor
transmitter with field-proven probes.
SITRANS LC300 Outputs
• 4 – 20 or 20 – 4 mA, 2-wire current loop circuit
• Isolated from the measurement circuit
• Current signalling according to NAMUR NE 43
Specifications
For a complete listing, see the SITRANS LC300 Instruction Manual.
Ambient/Operating Temperature
Process Conditions
Ambient temperature
– 40 to +85 °C (– 40 to +185 °F)
Pressure range
-1 to 35 bar g (-14.6 to 511 psi g)
Process temperature range
-40 to +200 °C (-40 to +392 °F)
Min. relative dielectric constant (
>
r
):1.5
Power
Short form
Long Form
Description
Units
LRV
Lower Range Value
value for 0 % (in pF)
4 mA
pF
pico Farads
10
-12
Farad
PV
Primary Variable
measured value
URV
Upper Range Value
value for 100% (in pF)
20 mA
Note:
SITRANS LC300 is to be used only in the manner outlined in this manual, otherwise
protection provided by the equipment may be impaired.
This product is intended for use in industrial areas. Operation of this equipment in a
residential area may cause interference to several frequency based communications.
• Supply voltage: 12-30 V DC any polarity, 2-wire current loop circuit, max. resistance
value 550
@ 24 V DC
Содержание sitrans lc 300
Страница 1: ...Capacitance Transmitters Quick Start Manual 03 2013 SITRANS LC300 SITRANS ...
Страница 2: ......