SITRANS F C MASSFLO
14
SFIDK.PS.028.R1.02
ENGLISH
Measurement of Mass flow [kg/s], volume flow [l/s], fraction [%], Brix, density [kg/m
3
], temperature [
°
C]
Current output
Current
0-20 mA or 4-20 mA
Load
<
800 ohm
Time constant
0-30 s adjustable
Digital output
Frequency
0-10 kHz, 50% duty cycle
Time constant
0-30 s adjustable
Active
24 V d.c., 30 mA,
1 K
Ω
≤
R
load
≤
10 K
Ω
, short-circuit-protected
Passive
3-30 V d.c., max. 110 mA,
1 K
Ω
≤
R
load
≤
10 K
Ω
Relay
Type
Change-over relay
Load
42 V / 2 A peak
Functions
Error level, error number, limit, direction
Digital input
11-30 V d.c.
Functionality
Start/hold/continue batch, zero-point adjust, reset totalizer 1/2, force output, freeze output
Galvanic isolation
All inputs and outputs are galvanically isolated, isolation voltage 500 volts
Cut-off
Low-flow
0-9.9% of maximum flow
Limit function
Mass flow, volume flow, fraction, density, sensor temperature
Totalizer
Two eight-digit counters for forward, net or reverse flow
Display
Background illumination with alphanumerical text, 3 × 20 characters to indicate flow rate, totalized
values, settings and faults.
Reverse flow indicated by negative sign
Zero-point adjustment
Manual via keypad or remote via digital input
Ambient temperature
Operation:
−
20
°
C
to +50
°
C
During storage:
−
40
°
C
to +70
°
C (Humidity max. 95%)
Communication
Prepared for client mounted add-on modules
Enclosure 19"
Material
Aluminium/steel (DIN 41494)
Rating
IP 20 to IEC 529 and DIN 40050 (1 m w.g. for 30 min.)
Mecanical load
18-1000 Hz random, 3.17G rms, in all directions, to IEC 68-2-36
Supply voltage
115/230 V a.c. +10% to
−
10%, 50-60 Hz
18-30 V d.c. or 20-30 V a.c.
Power consumption
230 V a.c.: 9 VA max.
24 V d.c.: 6 W I
N
= 250 mA, I
ST
= 2A (30 msec)
EMC performance
Emission
EN 50081-1 (Light industry)
Immunity
EN 50082-2 (Industry)
Ex-approval
[EEx ia] II C, DEMKO 03 ATEX 135251X
Maintenance
The flowmeter has a built-in error log/pending menu which should be inspected on a regular basis
Fuse
T 400 mA, T 250 V (IEC 127). Not replaceable by operator
Cable
Max. 300 m with cable code no.
FDK:
083H3005 or
FDK:
083H3006
C: max. 300 [pF/m]; L
c
/R
c
: max. 100 [
µ
H/
Ω
]
The total cable capacity have to be max. 200 nF
4.2 Signal converter MASS 6000
MASS 6000 19" IP 20/IP 65 version
4. Technical data
Terminals
81-82
83-84 & 89-90
85-86 & 87-88
U
o
[V]
16
15
15
I
o
[mA]
122
8
15
P
o
[W]
0.55
0.03
0.056
L
o
or
[mH]
2 or
10
10
L
o
/R
o
[
µ
H/
Ω
]
100
C
o
[nF]
200
200
200
Intrinsically safe