Installing/mounting
3.4 Mounting
Transducers
22
Compact Operating Instructions, 09/2018, A5E43390688-AB
①
Nipple welded to blind flange
②
Optional hazardous seal
Flange, gasket, and hardware supplied by customer.
Note
Tighten the flange bolts evenly in order to ensure a good seal between the mating flanges.
CAUTION
Potential performance degradation
Overtightening can cause performance degradation.
•
Do not tighten with excessive force.
①
Optional hazardous seal
④
Gasket (customer supplied)
②
Factory flanged transducer
⑤
Customer flange, flat face only
③
Bolt
⑥
Nut
Customer flanged standpipe. If a metal flange must be welded to pipe, refer to Liquid
Applications - Standpipes in the Transducer Applications Manual.