Commissioning
3 - 3
Siemens AG
SPR2-130.814.01
Page 3 of 10
SIREMOBIL Compact
Medical Solutions
Rev. 01 03.01
TD PS 24
System Manual
Enabling the Framegrabber Service Modes
3
•
Press the "OK/Enter" key on the control panel. The control panel displays the main
menu.
•
Press the "Erase/Cancel" key on the control panel 4 times.
•
Press the "Color Capture / Time" key on the control panel once. The display shows "Key
not supported...".
•
Press the "OK/Enter" key on the control panel.
•
Press the "Menu" key on the control panel.
•
Select the "Installation" menu line with the cursor key "
↑
" or "
↓
" and press the "OK/Enter"
key.
Programming the external exposure release
3
To enable the framegrabber to evaluate the print key on the Memoskop and SIREMOBIL
basic unit and to print an image, this function must be programmed.
•
The "Installation" menu is already selected. See "Enabling the Framegrabber Service
Modes".
•
If not, select the "INSTALLATION" menu line with the cursor key "
↑
" or "
↓
" and press the
"OK/Enter" key.
•
Select the "Exam ID Config" menu line with the cursor key "
↑
" or "
↓
" and press the
"OK/Enter" key.
•
Select the "Auto Date/Time" menu line with the cursor key "
↑
" or "
↓
" and press the
"OK/Enter" key.
•
Press the "Menu" key to select the main menu again.
If the installation menu cannot be selected, press the OK/Enter
key again and follow the instructions above under "Enabling the
Framegrabber Service Modes". The installation menu remains
enabled and selectable until the framegrabber is switched off.
NOTICE