92
43-TV-25-35 GLO Iss.4 Dec 06 UK
A5E01001767-04
Power up
•
FW Upgrade -
The Firmware Upgrade button allows new versions of the recorder
firmware to be loaded into the recorder.
•
Calibration -
Analogue In and CJC Calibration can be performed per slot (card).
•
Demo Traces
- A set of demo traces are available for demonstration or simulation pur-
•
Reset Setup -
Select this to clear the current setup. A Warning message will appear
before activating the reset.
•
Delete All Data
- This button will delete all data from the internal memory. A Warning
message will appear before data is deleted.
See “Delete All Data” on page 96.
When the configuration is complete select the Finish button to Commit, Discard or Commit
Later. Select the Back button to return to the previous menu.
Credits
(Configure > Setup > Edit Setup> General > Factory > Credits)
The Credit menu contains the Serial No. of the recorder, how may credits are available, dis-
plays the Options Code and lists what Firmware Options are currently active.
•
Serial No. -
read only and cannot be edited.
•
Credits -
read only and cannot be edited. It displays the number of Credits available to
activate additional firmware functionality.
•
Options Code -
The Options code is a unique coded number for this recorder that con-
tains information that enables certain functionality including the total credit value.
•
Options -
This displays how many credits are currently in use. Select this to produce a
list of Firmware options available and their credit values. Toggle each option On or Off
to enable the feature within the recorder. If there are not enough credits available. More
credits are available from your supplier. For a complete lists of all the firmware options
and their functionality see
Table 7.1, “Firmware Options,” on page 167
FW Upgrade
(Configure > Setup > Edit Setup> General > Factory > FW Upgrade
The Firmware Upgrade button allows new versions of the recorder firmware to be loaded
into the recorder via a USB key or a Compact Flash card. The file type required to upgrade
the Firmware has the file extension .xsu. Contact
SIEMENS
for more information. Insert the
Compact Flash or USB key with the latest revision of Firmware (.xsu file format) and press
the FW Upgrade button. The recorder will scan and check the files on the external device
with the current version running in the recorder. Any later version files on the external device
are copied into the recorder.
If there is no external media fitted or the file is corrupt a Firmware installation dialog will ap-
pear. If this occurs check the file is the correct format (.xsu). Secondly download the file to
another Compact flash card or USB key.
See “Storage Media Format” on page 120
for formatting information of Compact Flash
cards and USB keys. Compact Flash not available
for the
SIREC D200
.
To check the firmware upgrade has been successful, see the Status Menu,
Содержание SIREC D200
Страница 2: ......
Страница 176: ...170 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 Firmware Credit System ...
Страница 213: ...43 TV 25 35 GLO Iss 4 Dec 06 UK 207 A5E01001767 04 SIREC D400 Recorder 14 15 32 46 50 30 31 35 36 28 55 ...
Страница 214: ...208 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 52 11 23 33 39 5 6 RS485 USB host Ethernet 1 2 3 4 ...
Страница 218: ...212 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 SIREC D200 Recorder 16 5 3 1 4 2 24 49 9 45 19 6 49 ...
Страница 222: ...216 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 238: ...232 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 258: ...252 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 270: ...264 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 274: ...268 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 Modbus Function Codes ...
Страница 284: ...278 43 TV 25 35 GLO Iss 4 Dec 06 UK A5E01001767 04 ...
Страница 285: ......