POLYMOBIL III
Medical Solutions
Rev. 01
10.95
CS PS 24
Test with the unit in the operative condition
2
High-voltage test with the radiation detector
2
•
POLYMOBIL
ON
•
Connect radiation detector to the oscilloscope and place it in the beam path of the
POLYMOBIL III (see Fig. 1).
•
Set the slide switch on the radiation detector to measuring range 1 (1 point visible).
•
Set an SID of 40 cm and collimate onto the radiation detector with the light localizer
(Fig.1).
•
Trigger check exposures using the exposure data specified in the oscillograms (see
Fig.2).
The amplitude can be different for each detector.
Fig. 1
NOTICE