The daily total will be increased proportional to the amount of time the day was
lengthened.
Adjusting the Date
If the calendar is reset, the OCM-3 will adjust the log dates accordingly, taking into
account leap years and days per month.
EMULATION MODE
The flow calculation (P3/P4) can be checked for accuracy by using the emulation
parameter F1. The head is entered and the corresponding flow is displayed. This
function is useful when troubleshooting discrepancies between the OCM-3 calculation
and the expected flow.
Relays assigned to functions associated with the emulation parameter respond to the
emulated flows.
The mA output does not track the emulated flows when P28 (mA output emulation) =
0. However, if it is required to do so, then the emulator parameter should be set to 1.
RESET
The following resets can only be executed after satisfying the security access, F0.
Cold Start
If it is desired to reset all parameters, logs and totalizers to their factory setting, this is
done by forcing a cold start, F12.
Master Totalizer
If it is desired to reset the master totalizer (F2), this is done by parameter F11.
Data Log
If it is desired to reset the data log (F14), this is done by parameter F15.
Min/Max Log
If it is desired to reset the min/max log (F7), this is done by parameter F8.
7ML19985AB01
OCM III
35