Service Instructions
Multimobil 5C
Siemens Ltd. Med India
Version 5.0
Page 10 of 73
Copyright
©
SIEMENS LTD. All rights reserved. For internal use only
2.3.2 Disinfecting
As is commonly known, the ingredients in disinfectants
are hazardous to your health. Ensure that the room is
well ventilated when using any disinfectant. Follow
manufacturer’s instructions for the use of the product.
To disinfect surfaces, we recommend common liquid solutions of aldehyde- based or
ampholytic surfactant -based disinfectant.
Substituted phenol based or chlorine-releasing
disinfectant can weaken materials and are not
recommended. The same restrictions apply to undiluted
solutions with high alcohol content (e.g. for disinfecting
hands). Disinfectant sprays should generally not be
used, the spray can seep into the system, and safety
features can no longer be guaranteed (possible damage
to electrical components, formation of flammable
mixtures of air and the solution vapour).
As is known, some components of disinfectants are
detrimental to health. Their concentration in respiratory
air must not exceed a legally determined limit. We
recommend that appropriate directions on use issued by
the manufacture of these agents be strictly observed.
2.4 Protecting the system from fluids
Do not allow fluids to enter the system either during normal operation or during
cleaning and disinfections, as this may damage the system or cause a system
malfunction.
The Sheet metal cabinet of the control unit is not
watertight. The cleaning fluid should not seep into the
unit. Seepage of fluid into the unit can result in
malfunctioning of the electronic components
2.5 Routine
Checks
The Service personnel should perform the checks detailed in this section.
2.5.1 Daily checks
2.5.1.1
Prior to the Operation
Check that all safety related devices are functional and that the system is ready to
operate.
Please check:
-
The power cord. Do not use the unit if the power cord is damaged.
-
The additional conductive connection between the system and point of Potential
equalization, e.g. the patient table, must be established prior to the
examination.
CAUTION
NOTICE
WARNING
CAUTION