1 - 1
Siemens AG
SPL1-130.840.02
Page 1 of 2
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
General
1
Safety information and protective measures
1
•
Adhere to the safety measures described
Scope of applicability and regulations for the subsidiaries
1
Equivalent leakage current measurement
The equivalent leakage current must be measured where applicable under the require-
ments of DIN VDE 0751, Part 1.
Where DIN VDE 0751 does not apply, the subsidiaries should comply with the following
regulations (refer to ARTD - 002.731.17, as well as the safety technical regulations for
installation and maintenance).
The local national regulations apply primarily for the subsidiaries.
In the event that there are no existing local regulations, the following provisions should be
adhered to in the interest of the safety of customers, patients, employees and third parties
as well as the company.
Dangerous X-rax radiation and/or dangerous electrical voltage
during test and adjustment operations!
Disregarding safety precautions can result in death or serious
bodily injury.
The following points must be observed:
•
Disconnect the power cable when working on the system.
•
Ensure compliance with the general safety requirements when
working with the system under power.
•
Switch off the power prior to replacing modules or boards.
•
After all work has been completed and all covers have been
installed, perform the protective conductor test according to
ARTD-002.731.17...
•
The protective conductor resistance must not exceed 0.2 ohms.
•
When performing service work on the power-on module,
(replacing the power-on module or replacing the power cable),
the equivalent leakage current must be measured and recorded.
Checks and settings that need to be performed under X-ray radia-
tion are identified by the radiation warning symbol
.
When performing adjustments labeled as shown here, appro-
priate radiation protection measures must be taken (see ARTD,
Part 2).
WARNING