LITHOSTAR MODULARIS Error List
2 - 5
Siemens AG
SPL1-130.840.02
Page 5 of 14
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
Monitoring, E31
–
E34
2
E31
Error Text
:
Temperature sensor supplies incorrect data.
Possible
cause/action:
Temperature sensor o.k.?
Connection cable D3.X2 o.k.?
The temperature difference can be read out on the service PC.
E32
Error Text
:
The temperature at the shock wave head has exceeded the
maximum value.
Possible
cause/action:
Let the shock wave head cool down, but do not switch the
system off!
Check that the cooling unit is operating.
If M1 cooling pump is running pump head may need to be repla-
ced.
If the error recurs, check the function of the cooling unit.
E34
Error Text
:
Temperature cannot be determined: could not access the ADC
when converting the analog values from the temperature sen-
sor.
Voltage supply for the temperature sensors outside acceptable
range.
Possible
cause/action:
Connection cable D3.X2 o.k.?
Board D3?
Read out the temperature with the service PC.
Refer also to error 45.