Mechanical systems
10 - 13
Siemens-Elema AB
SPR8-215.898.01
Page 13 of 28
MOBILETT Plus / - E/ - M
Solna, Sweden
Rev. 05
01.02
SPS-UD
MOBILETT Plus and Plus M only
•
Remove the locking rings.
•
Remove the bearing rings.
•
Remove the shaft.
•
Lower the turn plate and remove it.
•
Check all springs for correct mounting and damage (Fig. 16).
•
Check that the dimension between the screw connection and the magnet housing is 49.7
mm. Adjust, if necessary.
A new or inspected turn plate should be installed in the reverse order:
•
Reinstall the turn plate.
•
Reinstall the shafts.
•
Reinstall the bearing rings.
•
Reinstall the locking rings.
•
Connect the ground cable and K68.
•
Reinstall the supporting wheels behind the tilt bar.
•
Reinstall the tilt bar.
•
Reattach the side covers and the rear cover.
•
Set switch S10 in OFF position to charge the capacitor bank.
•
Reattach the top cover.
•
Connect the unit to mains power supply outlet and switch it on.
•
Check the function of the turn plate by pressing one of the turn plate buttons and turning
the unit. Check also that the turn plate returns to the park position when the button is
pressed a second time.
Fig. 16
49.7 mm