Recalibration
7.4 Recalibration
BW500 and BW500/L
80
Operating Instructions, 12/2016, A5E33482052-AD
Span Calibration. Current Span
Setup test. Press ENTER to start
41900
For example: the current span count.
Note
End of span calibration. Remove the test weight and return to RUN.
7.4.7
Direct Span
Direct span entry (P368) is intended for use when replacing software or hardware, and when it is
not convenient to perform an initial span. A record of the last valid span count is required.
Access P368 and enter EDIT mode
P368-01 Direct Span Entry
Enter Span Count
E
0
Press the following keys in a given sequence:
P368-01 Direct Span Entry
Enter Span Count
V
41900
For example: the last valid span count.
Содержание Milltronics BW500
Страница 1: ...Operating Instructions Milltronics BW500 and BW500 L 12 2016 Edition Integrators ...
Страница 12: ...Table of contents BW500 and BW500 L 10 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 14: ...Safety Notes BW500 and BW500 L 12 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 16: ...The Manual BW500 and BW500 L 14 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 48: ......
Страница 68: ...Start Up 6 4 Load Cell Balancing BW500 and BW500 L 66 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 108: ...Operation 8 10 Totalization BW500 and BW500 L 106 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 153: ...Communications 11 7 Error Handling BW500 and BW500 L Operating Instructions 12 2016 A5E33482052 AD 151 ...
Страница 154: ......
Страница 198: ...Troubleshooting 13 2 Specifically BW500 and BW500 L 196 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 208: ...Appendix A 3 Technical support BW500 and BW500 L 206 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 218: ...Index BW500 and BW500 L 216 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 219: ......