Specifications
BW500 and BW500/L
18
Operating Instructions, 12/2016, A5E33482052-AD
Inputs
Load cell
0 ... 45 mV DC per load cell
Speed sensor
Pulse train 0 V low, 5 V high (TTL setting) or
10 ... 24 V high (HTL setting), 1 ... 3000 Hz or
open collector NPN switch or
relay dry contact
Auto zero
Dry contact from external device
mA
See optional mA I/O board
1)
Auxiliary
5 discrete inputs for external contacts (Low = 0 V, High =
10 ... 24 V), each programmable for either display scroll-
ing, totalizer 1 reset, zero, span, multispan, print, batch
reset, or PID function.
1)
Not available with BW500/L.
Outputs
mA
•
1 programmable 0/4 ... 20 mA, for rate, load, and
speed output
•
Optically isolated
•
0.1 % of 20 mA resolution
•
750
Ω load max
•
See optional mA I/O board
1)
Load cell
10 V DC compensated excitation for strain gauge type, 6
cells max, BW500 has 4 independent inputs, BW500/L
has 2. The maximum load, in either case, must not exceed
150 mA.
Speed sensor
12 V DC, 150 mA max. excitation
Remote totalizer 1
•
Contact closure 10 ... 300 ms duration
•
Solid state relay contact rated 30 V DC, 100 mA max.
•
Max. contact on-resistance = 36
Ω
•
Max. off-state leakage = 1 uA
Remote totalizer 2
•
Contact closure 10 ... 300 ms duration
•
Solid state relay contact rated 240 V AC/DC,
100 mA max.
•
Max. contact on-resistance = 36
Ω
•
Max. off-state leakage = 1 uA
Relay output
•
BW500
5 alarm/control relays, 1 form 'A' SPST relay contact per
relay, max. 5 A at 30 V DC / 250 V AC non-inductive; min.
10 mA at 5 V DC
•
BW500/L
2 alarm relays
1)
Not available with BW500/L.
Содержание Milltronics BW500
Страница 1: ...Operating Instructions Milltronics BW500 and BW500 L 12 2016 Edition Integrators ...
Страница 12: ...Table of contents BW500 and BW500 L 10 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 14: ...Safety Notes BW500 and BW500 L 12 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 16: ...The Manual BW500 and BW500 L 14 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 48: ......
Страница 68: ...Start Up 6 4 Load Cell Balancing BW500 and BW500 L 66 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 108: ...Operation 8 10 Totalization BW500 and BW500 L 106 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 153: ...Communications 11 7 Error Handling BW500 and BW500 L Operating Instructions 12 2016 A5E33482052 AD 151 ...
Страница 154: ......
Страница 198: ...Troubleshooting 13 2 Specifically BW500 and BW500 L 196 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 208: ...Appendix A 3 Technical support BW500 and BW500 L 206 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 218: ...Index BW500 and BW500 L 216 Operating Instructions 12 2016 A5E33482052 AD ...
Страница 219: ......