4 - 1
Siemens AG
RX57-029.061.01
Page 1 of 2
MEMOSKOP 2/3/50/100/SUB
Medical Engineering
Rev. 03
10.97
TD SD 34
Calling up the service menu
4
SIREMOBIL 2000
4
•
Switch the system "ON"
Calling up the memory and TV test images
4
•
Press and hold the following buttons on the SIREMOBIL control console in the
sequence shown:
1.
ATB (manual image release) button on the hand switch,
2.
Button (2/Fig.1) on the control console,
3.
and multiformat camera release button (3/Fig.1)
The "internal service menu" is displayed on the control console (Fig.1).
•
After approx. 1 second, release all the buttons.
Memory test images on SIREMOBIL 2000
4
•
Press the third button (soft key, 4/Fig. 1) on the control console and wait until a memory
test image is reconstructed (approx. 15 seconds).
•
Page through the memory test images by pressing the third soft key (4/Fig.1).
•
You can exit the service menu by pressing the same key combination as for calling up
the service menu.
LITHOSTAR Multiline
4
•
Call up the LITHOSTAR Multiline service menu.
Refer to "Setting the software parameters" RXL2-120.032.02...
Fig. 1
2
3
5
4