MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 2 of 6
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
3 - 2
Power up the MAMMOMAT Stand
1.
Connect the exposure lamp to the connector X759 located on the Master board
D750, see Fig. 2.
Mains voltage
3
The mains input can be either 1-phase or 2-phase, 50 or 60 Hz.
Electrical power to mains input:
3
Mains voltage
1-phase:
208, 230, 240 and 277 V ±10%
or
2-phase: 208,
230,
240, 277 and 400 V ±10%
Mains Frequency: 50 or 60 Hz
Maximum load:
10kVA (momentary) and 0.8 kVA (long time load).
Fuse:
25 A (external) 208 -> 400 V.
Do not use power from the MAMMOMAT stand.
Before the power for the MAMMOMAT is turned on, make sure,
that the cable plugs (left and right of the system) for the Magnifi-
cation table are not shorted.
Fig. 2
Connecting the exposure lamp
NOTE
WARNING
Magnification
table cable