Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 8 of 14
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
5 - 8
Cable connections
Installations with separate control console and separate generator
5
1.
Lay the connection cable from the separate control console in additional cable
ducts, alternatively underfloor, to the free cable entry of the generator, see Fig. 8.
2.
Lay the connection cable in a free U-bracket. Fold the shield braid over the
U-bracket and secure with a hose clamp as shown in Fig. 9.
Fig. 8
Connection cable from separate control console
Shielding the cable from the separate control console as
described below is an absolute demand to comply with the EMC
regulations.
Fig. 9
Shielding the control console cable
MAM00166
Connection cable from separate control console
NOTICE
M
A
M
00173
MA
M0
01
2
MA
M00
172