28
Cleaning
R630/R640; M440/M450
TD00-400.840.10.03.02
Siemens
02.08
CS SD SP/CO
Page 28 of 32
Medical Solutions
Preparatory work steps
6.1
•
De-energize workstation by removing the power plug from the mains supply.
•
The service technician must connect himself to the workstation with a suitable ground-
ing strap.
Electrostatic charging caused by the air flow of the vacuum cleaner must be avoided.
•
Remove the side panel.
•
Remove air duct(s).