Siemens
TD00-400.840.10.03.02
R630/R640; M440/M450
02.08
CS SD SP/CO
BIOS settings
19
Page 19 of 32
Medical Solutions
Changing back to the original BIOS settings
3.1
•
Reboot the computer
•
During the BIOS routine, press the F2 button
¹
The system will ask for the BIOS password.
•
Enter the password, according to the password list of your modality (to be found in the
knowledge database).
¹
The "BIOS Setup Utility" window appears.
Depending on your system, set the boot sequence back to the original settings as follows:
Setting the boot sequence on R640/R630
0
•
Select the "Boot" menu and then “Boot Device Priority”. Press Enter.
¹
The boot device list will appear.
•
According to the Help text on right side of the screen, set the the boot device back to the
original setting you noted previously.
•
Press the F10 button to save.
¹
You will be asked to "SAVE to CMOS and EXIT"
•
Approve with ’y’ and press Enter
¹
The computer will reboot automatically, using the new BIOS settings.
Setting the boot sequence on M440/M450
0
•
Select "Boot Options" in the Main menu. Press Enter.
•
Select "Boot Sequence". Press Enter.
¹
The boot sequence list will appear.
•
According to the Help text on right side of the screen, set the the boot device back to the
original setting you noted previously.
•
Press ESC twice to go back to the main menu. Go to the “Exit” menu and select “Save
and exit”. Press Enter.
¹
You will be asked to "SAVE to CMOS and EXIT".
•
Approve with ’y’ and press Enter.
¹
The computer will reboot automatically, using the new BIOS settings.