13
Guidance and manufacturer's declaration -
electromagnetic emissions
The HearCheck Screener is intended for use in the electro magnetic environment
specified below. The customer or the user of the HearCheck Screener should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The HearCheck Screener uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The HearCheck Screener is suitable for use
in all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Not applicable