Functions
7.13 Resistive Temperature Device (RTD) Calibration
FUG1010 IP65 NEMA 4X
136
Operating Instructions, 01/2013, A5E02951519-AC
5.
To enable numeric entry <Right Arrow>, then type in the reading of the reference
thermometer (e.g., 72.0).
6.
To recalibrate the RTD sensor <ENTER>. To verify the calibrated reading, go to the Dual
Path Flow menu [Diagnostic Data/Gas Data] menu to check the current RTD output.
Make sure that it coincides with the gas reading of the reference thermometer. Repeat for
the other RTD, if necessary.
Note
Factory Calibration provides an additional prompt after a new temperature is entered:
[Are you Sure? No Yes]. It is recommended that you use [User Cal] to avoid alteration of
preset factory calibration.
Ice Bath RTD Calibration
Use deionized water and ice mixture at 0°C (32°F) equilibrium for an ice bath. Ensure
temperature with a reference thermometer. Siemens can not assume responsibility for the
incorrect design, construction or operation of an Ice Bath.
NOTICE
Sensor Damage
If RTD sensor makes direct contact with ice during an ice bath calibration procedure the
sensor may be damaged and the calibration results will be incorrect.
Do not allow an RTD sensor to make direct contact with ice during an ice bath calibration
procedure.
To perform a 0°C (32°F) calibration:
1.
Immerse RTD sensor in deionized water and ice mixture. Stir the mixture constantly.
2.
In the [RTD Calibrate] menu move the highlight by the pressing <Up/Down Arrow> to the
RTD you want to calibrate (RTD 1 or RTD 2).
3.
To access the RTD option list press <Right Arrow>. Move the highlight to [User Cal] then
press <ENTER>. This triggers the pop-up window.
4.
After the RTD sensor reaches equilibrium at 0°C (32°F), press <ENTER> to recalibrate
the RTD sensor.
5.
To verify the calibrated reading, go to Dual Path Flow [Diagnostic Data/Gas Data] menu
to check the current RTD output. Make sure that it coincides with the gas reading of the
reference thermometer. Repeat for the other RTD, if necessary.
Содержание FUG1010 IP65 NEMA 4X
Страница 9: ...Table of contents FUG1010 IP65 NEMA 4X 8 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 23: ...Safety notes 2 6 Certificates FUG1010 IP65 NEMA 4X 22 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 93: ...Commissioning 6 3 Installation Menus FUG1010 IP65 NEMA 4X 92 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 143: ...Service and maintenance 9 4 Battery disposal FUG1010 IP65 NEMA 4X 142 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 165: ...Troubleshooting 10 5 Force Transmit FUG1010 IP65 NEMA 4X 164 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......
Страница 219: ......
Страница 220: ......
Страница 221: ......
Страница 222: ......
Страница 223: ......
Страница 224: ......
Страница 225: ......
Страница 226: ......
Страница 227: ......
Страница 228: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 232: ......
Страница 233: ......
Страница 234: ......
Страница 235: ......
Страница 236: ......
Страница 237: ...21614 C 1011NFPS 7 ...
Страница 238: ...1011NFPS 7 C 21614 ...
Страница 239: ...1011NFPS 7 C 21614 ...
Страница 240: ...21614 C 1011NFPS 7 ...
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ...21614 C ...
Страница 247: ......
Страница 248: ......
Страница 249: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 250: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 251: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 252: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 266: ...Glossary FUG1010 IP65 NEMA 4X 200 Operating Instructions 01 2013 A5E02951519 AC ...
Страница 270: ...Index FUG1010 IP65 NEMA 4X 204 Operating Instructions 01 2013 A5E02951519 AC ...