Appendix
Thickness Gauge
A6X30021999-06, 03/2012
7
Technical Data
Description
Data
Display Unit
128x32 LCD with backlight
Measuring Rate
5 Hz
Velocity Rate
1000 - 9999 m/s (3,280 to 32,805 ft/s)
Measuring Range
1.0 - 200 mm (0.04 to 7.99 inches)
Resolution
0.1 mm / 0.01 mm (0.01 in / 0.001 in)
Tolerance
0.65 - 9.99 mm ±0.04 mm
10 - 99.99 mm (±0.1% + 0.04) mm
100 - 250 mm ±0.03%
Power Supply
Two 1.5 V AAA dry cells
Size
61w x 108h x 28d mm
3
(2.4w x4.3h x
1.1d in)
Weigh
150g (5.3 oz) with batteries
Operating Temperature
-10
°
C to +50
°
C (14
°
F to 122°F)
Storage Temperature
-20
°
C to +60
°
C (-4
°
F to 140
°
F)
Probe Types and Measuring Range
Standard Probe:
PT-08 5.0 MHz
measuring range: 1.0 - 200 mm (.04 to
7.9 inches)
Battery Replacement/Disposal
Battery Replacement
The Thickness Gauge uses two 1.5 V AAA dry cells. To replace the batteries:
1.
Open the battery compartment cover by pressing the grooved section and sliding
the cover downwards.
2.
Remove the old batteries, insert two new 1.5 V AAA batteries and close the
battery cover.