background image

Manufactured by 
Sivantos GmbH
Henri-Dunant-Strasse 100 
91058 Erlangen
Germany
under Trademark License of Siemens AG.

www.bestsound-technology.de
www.bestsound-technology.com

Bezugsnummer:
Reference number:

D00095838

Ersetzt Ausgabe:
Replaces edition:

2014-03-21

Ausgabe/Edition:

2014-10-31

Life 5mi

SERVICE

Life 5mi

¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯  

Changing the Wireless System Address

Seite/Page:

16/16

Service Instructions on how to change the e2e 2.0 wireless system address

The information send by the hearing instrument is address coded. The standard address for all remote controls and 
hearing instruments systems is  "0 ". In case there are different users in a close neighbourhood cross communica-
tion can occur. To avoid this the complete system needs to be set to a new address. If you wear  two hearing inst-
ruments and use a remote control all devices have to be send for service.
Before changing the address the wireless functionality of the system should be validated by a wireless test.

Necessary measuring equipment:
PC, Noahlink, Connexx/HICOSS (Service software V 7.0 or higher), SIFIT (Service Software V 7.0 or higher)

1. Change the wireless address of a hearing instrument:

a) Activate Connexx and connect the hearing instrument
b) Detect instrument
c) Read out instrument
d) Choose folder " Fitting"
e) Press the button  "Configuration" --

>

 "Hearing Instrument"

f) Choose another Wireless Address

2. Change the wireless address of the Remote Control (TEK/miniTEK):

The wireless address of a remote control adapts automatically if it is connected.

3.

Change the wireless address of the Remote Control (ProPocket):
The service is not able to change the address.
A ProPocket with an alternative wireless address (Order number: 10 417 823) needs to be ordered from the 
manufacturer.

NOTE:
Please save the new data settings in your client’s documentation for further service cases.

Содержание 10666997

Страница 1: ...iedenwerden Dies giltauchfürdenSchalenwechsel Werkzeuge DiefürArbeitenundReparaturenanHörgeräten üblicherweiseverwendetenWerkzeugekommenzum Einsatz Service Manufactured by Sivantos GmbH Henri Dunant Strasse 100 91058 Erlangen Germany under Trademark License of Siemens AG www bestsound technology de www bestsound technology com Bezugsnummer Reference number D00095838 Ersetzt Ausgabe Replaces editio...

Страница 2: ...rk granite 10599330 grau grey 10599331 braun brown 10599333 schwarz black 10599334 silber silver 10599335 perlweiss pearl white 10599336 goldblond golden blonde 10599337 dunkelblond sandy brown 10599341 champagner dark champagne 10668324 Gehäuse Set CE ornament perlweiss Housing Set CE ornament perl white Batterielade Battery door 10676298 beige beige 10676308 granit granite 10676303 dunkles grani...

Страница 3: ... Beschreibung Description Seite Page 3 16 Manufactured by Sivantos GmbH Henri Dunant Strasse 100 91058 Erlangen Germany under Trademark License of Siemens AG www bestsound technology de www bestsound technology com 70 10426819 Achse D 0 6mm L 7 2mm Oberschale 1Stk Axle D 0 6mm L 7 2mm top housing 1pcs 10602389 Tragehaken gedämpft Earhook damped 10600047 Gehäusewerkzeug Housing tool ...

Страница 4: ...on 2014 03 21 Ausgabe Edition 2014 10 31 Life 5mi SERVICE Life 5mi Gerät Instrument Seite Page 4 16 Manufactured by Sivantos GmbH Henri Dunant Strasse 100 91058 Erlangen Germany under Trademark License of Siemens AG www bestsound technology de www bestsound technology com ...

Страница 5: ...e 5 16 Manufactured by Sivantos GmbH Henri Dunant Strasse 100 91058 Erlangen Germany under Trademark License of Siemens AG www bestsound technology de www bestsound technology com 10668332 Service Verstärker Service Amplifer 40 10666944 Batteriekontakt Battery Contact 50 10666943 Batteriekontakt Battery Contact 80 10296079 Wireless Spule 6 4 mm Wireless Coil 6 4 mm 110 10598794 Service Mikrofonset...

Страница 6: ...factured by Sivantos GmbH Henri Dunant Strasse 100 91058 Erlangen Germany under Trademark License of Siemens AG www bestsound technology de www bestsound technology com 120 10426806 Mikrofonlagerung Microphone Suspension 160 10598795 Service Hörerset Service Receiver Set 180 10296911 Hörerlagerung Receiver Suspension 190 10426810 Hörerschlauch Receiver Tube 200 10175174 Kontaktstück für Programmie...

Страница 7: ...y com 270 10296903 Akustische Verbindung Screw on connection 280 10175354 Schraube 0 8x3 5mm Screw 0 8x3 5mm 290 10172382 Service Label Service Label 10422386 Leitung 9 MM RD Solid Wire 9 MM RD 02766426 Leitung 12 MM BLK Solid Wire 12 MM BLK 02946668 Leitung 6 MM RD Solid Wire 6 MM RD 10297062 Leitung 6 MM BLK Solid Wire 6 MM BLK 10296127 Leitung 9 MM RD Litz Wire 9 MM RD 02756997 Leitung 8 MM RD ...

Страница 8: ... 2014 03 21 Ausgabe Edition 2014 10 31 Life 5mi SERVICE Life 5mi Verstärker Amplifier Seite Page 8 16 Manufactured by Sivantos GmbH Henri Dunant Strasse 100 91058 Erlangen Germany under Trademark License of Siemens AG www bestsound technology de www bestsound technology com ...

Страница 9: ...s 10419704 Service Indikator Clips blau 100 Stk Service Indicator Clips blue 100 pcs 10419685 Indikator Clips rot blau je 2 Stk Indicator Clip red blue each 2 pcs 10296944 LifeTube Refill 1L LifeTube Refill 1L 10296945 LifeTube Refill 1R LifeTube Refill 1R 10296946 LifeTube Refill 2L LifeTube Refill 2L 10296947 LifeTube Refill 2R LifeTube Refill 2R 10296948 LifeTube Refill 3L LifeTube Refill 3L 10...

Страница 10: ...e Shaper Refill 4 10054970 Life Cleaning Wire Refill Life Cleaning Wire Refill 10054711 LifeTube Schablone LifeTube Selector 10296968 LifeTube Fitting Set LifeTube Fitting Set 10673779 Semi Open Tip Set von 10 Stück Semi Open Tip set of 10 pieces 10296987 Open Tip 4MM Set von 10 Stück Open Tip 4MM Set of 10 pieces 10296988 Open Tip 6MM Set von 10 Stück Open Tip 6MM Set of 10 pieces 10296989 Open T...

Страница 11: ... V7 0 oder höher 1 Niemals das Mikrofonpaar Sach Nr 10 598 794 trennen Immer komplettes Mikrofonpaar ersetzen Wenn ein Mikrofon ausgefallen ist müssen beide Mikrofone aus gewechselt werden 2 Zur Überprüfung und Abgleich des vorderen Mic1 und hinteren Mic2 Mikrofons ist folgendes zu beachten a Eingangspegel 60 dB FONIX 6500 b Die Verstärkungseinstellung wird durch HICOSS automatisch eingestellt Mic...

Страница 12: ...itte die geräte spezifischen Kissen in die SCAL Testvorrichtung Das Kissen für den Gehäusedeckel be sitzt zwei Löcher für die Haltestifte Legen Sie bitte eine Batterie in das Gerät ein oder schließen Sie das Hörgerät über einen Batterie Adapter an eine konstante Spannungsquelle 1 3V an Verbinden Sie den Geräteausgang Haken mit dem in der Testvorrichtung liegenden Schlauchende Stecken Sie das Progr...

Страница 13: ...est system HICOSS Service Software V7 0 or higher SIFIT V7 0 or higher 1 Never separate the microphone pair Order No 10 598 794 The complete pair must be replaced If one microphone fails both microphones must be replaced 2 Front Mic1 and Rear Mic2 test and adjustment a Input sound pressure level Fonix 60 dB b The gain is adjusted automatically through HICOSS Calibration Mic1 F11 Mic2 F12 c Positio...

Страница 14: ...following SCAL components SCAL Basic Jig Order No 7222529 Silicon cus hion Life Order No 10 417 430 Please insert the device specific cushions into the SCAL jig Top case cushion has two holes for inserting the fixing pins Please insert battery into the device or connect it via battery pill to a constant voltage drive 1 3V Connect ear hook of the device with the end of the jigs internal tube Connec...

Страница 15: ...tzt Ausgabe Replaces edition 2014 03 21 Ausgabe Edition 2014 10 31 Life 5mi SERVICE Life 5mi Replacing the Twin Microphone Seite Page 15 16 Zur genauen Positionierung der Mikrofone wird wie in der Zeichnung gezeigt empfohlen auf das Gitter der FO NIX Box ein Fadenkreuz einzuzeichnen To ensure the proper positioning of the microphones it is recommended to draw a reticule on the grid of a FONIX Box ...

Страница 16: ...l all devices have to be send for service Before changing the address the wireless functionality of the system should be validated by a wireless test Necessary measuring equipment PC Noahlink Connexx HICOSS Service software V 7 0 or higher SIFIT Service Software V 7 0 or higher 1 Change the wireless address of a hearing instrument a Activate Connexx and connect the hearing instrument b Detect inst...

Отзывы: