8014084/ZWT7/V3-1/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP
Subject to change without notice
4
OPERATION
4.4.3
Calibration gas, actual value (CAL.C)
1
Holds the analog output.
2
Displays the current O
2
concentration measured.
Fig. 19: Calibration gas, actual value display
4.4.4
Signal intensity (SIL)
1
Compares the current signal intensity of the laser on the receiver against the signal level
from the factory calibration.
2
The signal intensity serves to determine contamination on the optics. Important: The
laser signal can be amplified so that values above 100% are possible.
Fig. 20: Signal intensity display
4.4.5
Display of active and undeleted errors (ERR)
This menu displays all active error messages.
describes how the errors are read out
and deleted. The display shows
E 0
when all errors have been deleted. Significance of error
numbers,
Fig. 21: Displaying all errors currently existing