A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | S710 E
Subject to change without notice
2
ENGLISH
2.4
Setting the measuring range
Fig. 4: Operating and display elements
1
Press “ESC” as often as required until you reach the main menu.
2
Press “.” (decimal point) to extend the menu selection.
3
Press key “6” to go to the “settings” menu.
4
A warning is displayed: Read the warning and confirm with “Enter”.
5
Press key “2” to go to the “measurement” menu.
ModularSystem
Function
Service
Alarm
7
8
9
4
5
6
1
2
3
0
Esc
Help
Enter
Main menu
1 measuring display
2 device status
3 control
4 calibration
5 maintenance signal
Enter digit
No messages
CO 12 mg/m3
←
standard functions
←
←
←
←
←
operation information
←
status messages
←
measured values (alternating)
Main menu
1 measuring display
2 device status
3 control
4 calibration
5 maintenance signal
6 settings
7 service
Enter digit
No messages
CO 12 mg/m3
←
“settings”