9
Replacing the system plug
NOTICE
Enclosure rating IP65 only applies if the safety laser scanner is closed and the system
plug is mounted.
b
Mount the system plug and the cover plate.
b
Close each M12 plug connector on the safety laser scanner using a male cable
connector or a protective cap.
°
Tightening torque for plug connector: 0.4 Nm … 0.6 Nm.
°
Tightening torque for protective caps: 0.6 Nm … 0.7 Nm.
b
Mount the optics cover.
NOTICE
Plug in the system plug carefully.
Do not force it.
The contacts may break off or bend if too much force is used.
Replacing the system plug:
Tool required:
•
TX20 Torx wrench
Approach
1.
Make sure that the environment is clean and clear of fog, moisture, and dust.
2.
Disconnect the connecting cables the system plug.
3.
If necessary: move the safety laser scanner to a clean location.
4.
Unscrew screws in the defective system plug and remove the system plug from the
safety laser scanner.
5.
Carefully insert the new system plug into the safety laser scanner.
6.
Screw in the system plug using the captive screws. Tightening torque: 2.25 Nm to
2.75 Nm.
7.
Reconnect the connecting cables to the system plug.
8.
Configure the safety laser scanner.
9.
Perform commissioning again, taking particular care to conduct all of the thorough
checks described. For more information, see the operating instructions.
MOUNTING INSTRUCTIONS
16
M O U N T I N G I N S T R U C T I O N S | microScan3 – EtherNet/IP™, microScan3 – EFI-pro, microScan3 – PROFINET (M12)
8020210/14BK/2019-07-04 | SICK
Subject to change without notice