5.4
Adjustment
Menu 2: Adjustment
5.4.1
Adjustment/manual
Menu 2.1: Adjustment/manual
5.4.1.1
Zero point and span point
Menu 2.1.1: Adjustment/manual/Zero point
Menu 2.1.2: Adjustment/manual/Span point
Manually adjust zero point and/or span point (recalculate the correction factor) of
individual components within this menu item.
2
see “Adjustment/manual”, page 44
see “Adjustment/automatic”, page 45
see “Adjustment/Parameter”, page 45
Status: Measuring
Back
Enter
Adjustment
/Adj
1 manual
2 automat
3 Parameter
2.1
see “Zero point and span point”, page 44
see “Zero point and span point”, page 44
Status: Measuring
Back
Enter
manual
/Adj/man
1 Zero point
2 Span point
2.1.1
1
Set the maintenance signal (
see “Maintenance/maintenance signal”,
).
2
Lead the zero and/or reference medium into the cell (“manually”).
3
Wait for the end of the run-in time, (If necessary, exit the menu with
“
Back
” to review the run-in behavior in the line graph,
4
Select the component.
5
Press “
Set
”: The measured value is set to the nominal concentration. If
the deviation is too large (parameter settings
→
SOPAS ET), the
MCS300P Ex switches to classification “
Maintenance request
” (
“Status and classification”, page 36
6
Press “
Back
” to exit the menu.
7
Reset the maintenance signal.
(Span point menu accordingly)