O P E R A T I N G I N S T R U C T I O N S | MCS300P Ex
Subject to change without notice
5
MENUS
5.3.2
Maintenance/operating states
Menu 1.2: Maintenance/Operating states
5.3.2.1
System stop
Menu 1.2.1: Maintenance/Operating states/System stop
1.2
The operating states are switched on in this menu.
Status: Measuring
Back
Start
Operating states
/Maint/Opstate
1 System stop
2 Measuring
3 Initialize
▸
System stop (
▸
Switch the MCS300P Ex to measuring operation.
▸
New start (reset) of the MCS300P Ex.
********
Switch off the measuring function and (If available) the gas
sampling peripherals (valves, pumps).
The device function is maintained (e.g., heater operation will
continue).
Operating state:
“Maintenance”.
Status: System stop
Back
System stop
●
Display: Row of blinking asterisks.
▸
To exit the menu: Press the “
Back
” button.
▸
Then switch measuring operation on again:
- Press “
Measuring
”.
- Switch the maintenance signal off (