27
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
MAINTENANCE
3
4
Open the cell cover (
see Chapter “Working on the cell”, page 14
Fig. 24: Ejector block, hose connections
5
Disconnect the cable connection of the oxygen sensor:
▸
Disconnect the plug connection on the device cabinet.
▸
Loosen the fastening clip with a Phillips screwdriver.
6
Disconnect hose connections 2 and 3 with a jaw wrench (
, item numbers 3 and 4).
Fig. 25: Hose slots
1 Ejector block
2 Oxygen sensor
3 Hose connection 2
4 Hose connection 3
5 Hose connection to cell inlet filter
6 Exhaust gas line
The hose connections are marked with numbers on the ejector block and on the slots to
the left of the cell.