13.2.3
Housing
Table 15: Design
Design
Design
1 x stand-alone enclosure
Material, general
Steel plate, aluminium cast
Dimensions
Installation
Upright
Weight
Approx. 200 kg
Materials with media contact
•
Stainless steel 1.4571
•
PTFE
•
Aluminium (coated)
Degree of protection
IP 54
Impact resistance
IK08
13.2.4
Interfaces and protocols
Table 16: Interfaces and protocols
Operation and interfaces
Operation
Via display or Google Chrome browser with SOPASair
software, several operating levels, password-protected
Display and input
Foiled color display with touchscreen
Analog inputs/outputs
Optional
Digital inputs/outputs
Optional
Data interface
1 x Modbus TCP/IP
Profibus
Configurable
Profinet
Configurable
Remote support
SICK MPR (optional)
PC operation
Browser Google Chrome with SOPASair via Ethernet
13.2.5
Power supply
Table 17: Voltage supply
Voltage supply
Power input
•
Analyzer
•
Heated sample gas line
•
Gas sampling unit
•
Heated probe tube
Power input
•
Approx. 1000 VA
•
Approx. 95 VA/m
•
Approx. 450 VA
•
Approx. 450 VA
Table 18: Optional interfaces
Interfaces (optional)
Digital outputs
4 outputs, 24 V, 0.5 A
Digital inputs
Electrically isolated, 24 V, 0.3 A
Table 19: Cable glands
Cable glands
Hose bundle line
M40x1.5 D22 -32 IP68 PA-GR
Main power supply
M32x1.5 D18 -25 IP68 PA-GR
UPS power supply
M20x1.5 D10 -14 IP68 PA-GR
13
TECHNICAL DATA
64
O P E R A T I N G I N S T R U C T I O N S | MCS200HW
8021889/1D1T/V3-1/2021-09 | SICK
Subject to change without notice