5.7.9
Configuration
Main Menu: Configuration
5.7.9.1
Password for the configuration level
A password has to be entered to edit the menus “Signal output” and “Graphic output”.
The password input is not displayed on the screen.
5.7.9.2
Signal output
Main Menu:
Configuration: Signal output
This menu option is used to define the:
• analog signal outputs and the
• Measurement ranges
Max. 32 outputs can be activated.
The menu option is identical to
Specialist: System edit: Signal output
Definition of password:
Settings
Channel
Number of the analog output.
Mark (
)
Analog output active (
)/not active.
Name
Name of component derived from a list which is based on
Specialist: System edit:
Results
.
Display list by pressing <ENTER>
Unit
Unit of component. Displayed automatically from
Specialist: System edit: Results
Measurement
ranges
Switching of the measurement range is programmed here.
Measurement range 2 is the larger measurement range and relevant for the plausibil-
ity check for zero setting- and for calibration, with the drifts etc.
The measurement ranges are automatically switched over (this is also true for the
graphic output). Switching over can be interrogated in the menu
Specialist: System
edit: Zero/Calib/Purge programs/other programs
(Variable RHi).
Hysteresis is 5%.
If a component has been defined more than once, then the 1
st
definition is relevant
for the plausibility check.
Relays
Number: Number of the relay
The relay “Number” is activated (deactivated) when the measurement range 2 (1) is
active (the relay indicates that the analog output switched to another output range).
0: No relay to be assigned
- Number: Inverted setting
Current
0..20 mA
4..20 mA
Содержание MCS100E
Страница 1: ...O PERATING INSTRU CTIO NS MCS100E IR Analyzer Title Page Commissioning Operation Maintenance...
Страница 10: ...Contents 10 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG...
Страница 98: ...Operation 98 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...
Страница 112: ...Maintenance 112 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...