9.1.3
Menus
9.1.3.1
Runtime Parameters
Serial buffer size
Default setting of serial buffer: 1024 bytes.
Maximum size that can be set: 64 Kbytes.
If the buffer memory reserved is exceeded during measurement, then MCS100E:
• Generates an acoustic warning signal.
• The message “Faulty sequence” is presented in the menu option “Messages”.
Data saving period
The menu option “Data saving period” is used to define a time after which the data files
created are saved periodically on the silicon disk/hard disk.
As soon as this period has elapsed (min. 60 sec) the program saves the data, even if the
maximum file size was not yet reached.
If the maximum file size is reached before the end of the selected data saving period, then
the program stores the data file at an earlier time.
The default setting for the data saving period is 3000 sec.
Number of sortable files
When calling-up the menu for listing the data files, the names of the files stored in a
directory are read into a buffer and presented on the screen.
The menu option “Number of sortable files” can be used to specify the maximum number
of files that can be organized in the buffer. Default setting is 127 files. The buffer size can
be varied within a range between 63 to 511.
Valid places in number representation
Number of significant places for numeric output of the measured values in bar graph
representation (in time history the display shows always the complete sequence of digits
following the decimal point).
Default setting: 6 significant places.
Possible setting: 1 ... 10 significant places.
This setting can be changed in the operator menu in the graphic representation
Main
Menu: Configuration: Graphic output
.
Password (definition of passwords)
Four passwords per level can be defined.
If no password is entered: Skip the password prompt with <ENTER>.
Passwords
Remark
1 ..4
– Valid for:
- “Configuration menu” level
– The pre-set password is: 88709
5 ..8
– Valid for:
- “Specialist menu” and
- quitting the program
– The pre-set password is only available for
trained personnel.
Even if just one password is missing, there is
no
password protection.
For password protection it is therefore necessary to enter all 8 passwords
(even if these are identical).
For the specialist menu: Contact SICK Service.
Содержание MCS100E
Страница 1: ...O PERATING INSTRU CTIO NS MCS100E IR Analyzer Title Page Commissioning Operation Maintenance...
Страница 10: ...Contents 10 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG...
Страница 98: ...Operation 98 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...
Страница 112: ...Maintenance 112 MCS100E Operating Instructions 8009504 VYWA7 V3 1 2018 01 SICK AG Subject to change without notice...