Table 5: Procedure phases during manual adjustment
Actual state
Internal function
Stop
Function idle
Test gas
Wait for flush time to elapse (after switching to test gas)
Measuring
Determine measured values with the test gas
Calculate
Calculate mean value from measuring time, calculate devia‐
tion from setpoint, adapt adjustment
Sample gas
Wait for flush time to elapse (after switching to sample gas)
Table 6: Information in menu
Manual adjustment
during the procedure
Identifier
Significance
Actual state
Part of the procedure running (
)
Meas. value
Current measured value of component
Actual countdown timer (SCCDGi)
Name of countdown timer running
Remaining time
Remaining time of countdown timer running
5
ADJUSTMENT FUNCTIONS
42
T E C H N I C A L I N F O R M A T I O N | MARSIC200
8017324/15A2/V6-0/2019-10 | SICK
Subject to change without notice