å
= the test gas is used for reference point validations with adjustment
cuvette.
Simultaneous use for reference point validations is not possible.
æ
= the test gas is used for reference point adjustments with adjustment
cuvette.
Simultaneous use for reference point adjustments is not possible.
ç
= the test gas is used for linearity adjustments.
Simultaneous use for other adjustments is not possible.
1
Tag format: SiMVj (i = Sensor module number, j = measured value number in the Sensor module)
2
The longest flush time is effective when the test gas is used simultaneously for several components.
NOTE
The settings are first effective after
Save
has been selected.
5.2
Performing a manual adjustment
Function
“Manual adjustment” means a single adjustment or validation procedure is selected
and manually started.
NOTE
The standard setting with checkbox “Automatic” selected (
)
means the test gas is fed automatically via solenoid valves controlled by the MAR‐
SIC200 digital outputs. The controlling output is defined in the respective test gas set‐
ting (“Action on start”
see "Configuring test gases (Test Gas Table)", page 39
). The test
gas can also be fed manually.
Procedure
b
Inform connected locations on the impending interruption in measuring operation.
1.
Call up
BCU
/
Maintenance
/
Manual adjust
.
2.
Select the measuring component for which the procedure is to be applicable
([ << ] [ >> ]).
3.
Select the desired function ([ << ] [ >> ]).
✓
The
Start
button is displayed when a suitable test gas setting exists for the mea‐
suring component and function combination.
4.
Select
Start
.
With manual test gas feed
With automatic test gas feed
✓
Actual state =
Test gas
5 Feed the suitable test gas into the sam‐
ple gas inlet of the MARSIC200.
6 Wait until
Sample gas
is displayed as
actual state.
7 Now feed sample gas into the sample
inlet again.
1
8 Wait until
Stop
is displayed as actual
state.
✓
The manual adjustment has com‐
pleted.
✓
The automatic procedure starts.
Actual state = part of the procedure
currently running (
)
5 Wait until
Stop
is displayed as actual
state.
✓
The manual adjustment has com‐
pleted.
1
Alternative (when a further manual adjustment is to follow): The test gas for the next manual adjust
ADJUSTMENT FUNCTIONS
5
8017324/15A2/V6-0/2019-10 | SICK