NOTE
Pay attention to the function indicator of the terminals!
Example: 1-X3:9 “Standby” to XD62:2 “Standby”.
Line
Connection
External power supply on distribution unit
Distribution unit:
see "Electrical installation of distribution
Power supply from distribution unit to analyzer
Distribution unit:
see "Electrical installation of distribution
Analyzer:
Status signals between analyzer and distribution unit
Analyzer:
Distribution unit:
see "Electrical installation of distribution
External analog and digital signals, Ethernet to analyzer
Analyzer:
2.7
Installing the sampling probe
The sampling probe is system-specific: For information on the installation of the sam‐
pling probe, see the enclosed Operating Instructions of the sampling probe.
NOTICE
Risk of soiling the measuring system
b
First install the gas sampling system on the exhaust duct just before the analyzer
is switched on.
b
Switch the instrument air feed on immediately after installing the sampling tube.
Installation
b
Install the sampling probe in accordance with the specifications in the Operating
Instructions of the sampling probe.
w
Fit the sampling tube with the probe tip tilted down about 10°.
min.70
10°
~
~30
INSTALLATION
2
8017324/15A2/V6-0/2019-10 | SICK
T E C H N I C A L I N F O R M A T I O N | MARSIC200
21