background image

6

   

ANNEX 

34  

Operating Instructions | InspectorP Rack Fine Positioning 

8024534 /   / 2019-08-19 

Parameter 

Description 

Use 

 

sensitivity can be improved by speci-

fying the minor axis length. 

Ethernet mode 

Set communication protocol for Ether-

net interface. 

Adjust setting to the communication 

protocol of the control used.  

Gain factor 

Manual adjustment of image bright-

ness. 

Adjust the image brightness until the 

positioning mark is as visible as pos-

sible. 

Restart device 

Device is restarted. 

Restart the device in the event of er-

rors. 

Display saved image 

Display the currently saved image in 

SOPASair.  

The saved image is displayed and 

current continuous measurement is 

switched off. 

Recalculate saved image 

Reprocess and evaluate the image 

currently saved in the device. 

The saved image is processed again 

with the adapted settings of the algo-

rithm. 

Light circle on dark back-

ground (reflector) 

Set polarity of the position mark (light 

circle on dark background or dark cir-

cle on light background). 

Holes typically appear dark on light 

backgrounds. Reflectors appear light 

on dark backgrounds. 

Inverted X/Y axis 

Exchange output values for X-axis and 

Y-axis.  

Use this parameter, for example, if 

the device is installed upside down in 

the system.  

FTP server IP address 

Set IP address of the FTP server.  

Define target address for image 

transmission. 

PLC IP address 

Set IP address of control.  

Define target address for data output 

to the control. 

IP port InspectorP for TCP 

(PLC is client) 

Set IP ports for the device. 

Define IP ports if the device is in the 

TCP server network. 

Compact mode 

Set compact display. 

Function currently not supported. 

Continuous measurement 

Activate device. Image recording is 

started continuously and each image 

is evaluated. 

Record and evaluate images continu-

ously during commissioning. During 

actual operation, continuous meas-

urement is activated by the control 

via a command as soon as the auto-

mated storage and retrieval system is 

in front of the rack and the position-

ing mark is in the field of view of the 

device.  

Circle diameter 

Define circle diameter of the position-

ing mark to be searched for.  

Optimize performance and position-

ing accuracy. 

Minimum acceptance score 

Minimum measured value which 

must be reached to detect a hole or a 

reflector as a positioning mark. 

The lower the measured value, the 

more elements are detected as the 

positioning mark. This can be useful 

to allow for detection of circles with 

unclear outlines (e.g. caused by corro-

sion). A value which is too low leads 

to error messages.  

 

Содержание InspectorP63 Flex C-mount Series

Страница 1: ...OPERATING INSTRUCTIONS InspectorP Rack Fine Positioning 2D vision ...

Страница 2: ... may also be trademarks and are only used here for identification purposes Legal information This work is protected by copyright The associated rights are reserved by SICK AG Re production of this document or parts of this document is only permissible within the limits of the legal provisions of copyright law Any modification abridgment or transla tion of this document is prohibited without the ex...

Страница 3: ...and use 10 3 3 Principle of operation 11 3 4 Single deep applications 12 3 5 Single and double deep applications 12 3 6 Positioning mark 12 4 Installation 14 4 1 Software installation 14 4 1 1 SICK AppManager 14 4 1 2 SOPASair web server 14 4 1 3 Saving parameter set 15 4 1 4 Installing firmware updates 16 4 2 Mounting 16 4 2 1 Mounting procedure 16 4 2 2 Mounting device 16 4 3 Electrical installa...

Страница 4: ...brightness 23 5 3 8 Setting up Ethernet interface 24 5 3 9 Additional parameters 24 5 3 10 Completing configuration 25 5 4 Specification of the interface and interaction with the control 25 5 4 1 Communication process 25 5 4 2 Interfaces 25 5 4 3 Commands from the control to the device 26 5 4 4 Answer from the device to the control 29 5 4 5 Control via digital inputs and digital outputs 30 6 Annex...

Страница 5: ...y these operating instructions should be handed over with it These operating instructions do not provide information on operating the machine or system in which the device is integrated For information about this refer to the oper ating instructions of the specific machine Related publications The following related publications must be read Part number language Publication 8019946 English 8019945 ...

Страница 6: ...ient and trouble free operation 1 3 Further information NOTE Other documentation for the device can be found on the online product page at www sick com InspectorP_Rack_Fine_Positioning There additional information has been provided depending on the product such as Model specific online data sheets for device variants containing technical data dimensional drawing and specification diagrams These de...

Страница 7: ... with the respec tive applicable safety standards for machines The device must not be used in explosion hazardous or corrosive areas or under ex treme ambient conditions Any use of accessories not specifically approved by SICK AG is at your own risk WARNING Danger due to improper use Any improper use can result in dangerous situations Therefore observe the following information Device should be us...

Страница 8: ...sult in considerable personal injury and material damage All work must only ever be carried out by the stipulated persons This product documentation refers to the following qualification requirements for the various activities associated with the device Instructed personnel have been briefed by the operating entity about the tasks as signed to them and about potential dangers arising from improper...

Страница 9: ... knowledge of the design and setup of the described connections and interfaces Basic knowledge of data transmission Knowledge of the programming of image pro cessing systems and network components Operation of the device for the spe cific application Knowledge of the operation and control of the de vices in their particular application Knowledge of the software and hardware environ ment for the pa...

Страница 10: ...the focus adjustment screw The la bel can be stuck over the access opening for the screw on the top of the device af ter adjustment 1 Printed safety notes multilingual 8022655 Brief information and safety notes for de vices of series ICR6xx and V2D6xx 3 2 Function and use Temperature load and steel construction dependent influencing variables generate varying geometric conditions at transfer point...

Страница 11: ...lf is twice as deep The device is able to handle both applications Depending on the application type of positioning mark rack depth there are differences in system design such as the se lection of the lens focal length and the use of lighting filters The positioning mark hole or reflector is typically placed in the center of the camera image and defined as the target region during alignment and ra...

Страница 12: ...on parameter Applications with holes Holes as positioning marks can be used both for single deep applications as well as combined applications with single and double deep racks Applications with reflectors Round reflectors as positioning marks can only be used for single deep applications NOTE Using the following reflectors is recommended PL22 1 part number 1003546 PL22 2 part number 1003621 PL22 ...

Страница 13: ...of polarizing filter kit part number 2088229 Polarizing filter disk 2 Turn optics cover counterclockwise 3 Remove optics cover 4 Turn spacer ring counterclockwise 5 Remove spacer ring 6 Clip on polarizing filter disk making sure the alignment is correct Centering ring downwards recesses over retaining clips 7 Put on optics cover 8 Turn optics cover clockwise Figure 3 Polarizing filter mounting ...

Страница 14: ...es Installing SICK AppManager To install SICK AppManager open the installation file exe and follow the instruc tions on the screen SICK AppManager user interface 1 Device window window name changes depending on the connected device 2 Device Search window 3 Firmware window 4 1 2 SOPASair web server The device is set up and parameterized using the SOPASair operating software You can call up the oper...

Страница 15: ...necessary Saving on the PC back up 1 In the SICK AppManager in the Device window click on the Backup button The Backup Apps and Data window opens 2 Fill out the Name Version and Filename fields 3 Select the path for data back up 4 Select all available options 5 To run the back up click on the Backup button in the Backup Apps and Data win dow Saving on a microSD memory card cloning NOTICE Risk of d...

Страница 16: ...sioning The optics are already pre focused at a distance which is suitable for most applica tions It might be a good idea to optimize the focus position in rare cases see Figure 8 Manual focus screw page 22 The procedure for mounting the device is divided into the following steps Mount device with a view to the positioning mark Connect device to interfaces and supply voltage Align device with aimi...

Страница 17: ...t interface Furthermore the device transfers the deviation between the actual and the target position of the automated storage and retrieval system to the control via the Ethernet interface In doing so the device precisely forwards the deviation between the actual position and the target position to the control using X and Y values Based on these values the control initiates position correction of...

Страница 18: ...gital outputs Pin 10 Sensor 1 digital input IN 1 switch continuous measurement on and off Pin 15 Sensor 2 digital input IN 2 activate and deactivate double deep mode Pin 13 IN OUT 3 direction specification up Pin 14 IN OUT 4 direction specification down Pin 16 IN OUT 5 direction specification right Pin 17 IN OUT 6 direction specification left For further information on digital inputs and digital o...

Страница 19: ...k 3 Select the desired device The Device window shows the application currently loaded on the device 4 Click on the Edit IP Adress button The Edit IP Adress window opens 5 Select the Use manual IP settings option 6 Enter an IP address 7 Confirm changes with OK 5 2 Establishing a connection PROFINET In order to use the PROFINET communication protocol the local device setting must be in the Profinet...

Страница 20: ...r level click on the button with the currently set user level at the top right in the user interface The following user levels are available User levels Password Application Run No password required Observe the system during operation Switch continuous measurement on and off Activate and deactivate double deep applications Switch image output on and off Operator No password required Additional set...

Страница 21: ...he device After the device has been roughly aligned during mounting fine adjustment is done us ing the SOPASair operating software Prerequisites The device has been mounted at the same height as the positioning mark The device is aligned vertically to the rack with a view to the positioning mark The Operator or Maintenance user level is set 1 Open the Operation window in SOPASair 2 Activate the Co...

Страница 22: ...n the front screen of the device and the rack with the positioning mark 3 Make the following calculation Measured distance between front screen and rack 40 mm calculated working distance Example 300 mm 40 mm 340 mm 4 Enter the calculated working distance in the Working distance area into the Work ing distance single deep mm input mask With double deep applications enter the calculated working dist...

Страница 23: ...ep rack is visible in the image Test fine adjustment and calibration in single deep applications 5 3 6 Adjusting target region The size of the target region can be adjusted The closer the set size is to the actual circle diameter of the positioning mark the more exact the positioning of the auto mated storage and retrieval system is However this means that more position adjust ments have to be mad...

Страница 24: ...er in Double deep mode to a value which offers sufficient image brightness for positioning marks in double deep racks Do not set values greater than 5 to prevent increased image noise If higher values are nec essary adjust the illumination time accordingly 7 Check the settings in Single deep mode again If necessary reduce the Gain fac tor parameter in Single deep mode 8 Activate the Perform auto g...

Страница 25: ...smitted in micrometers Determining the average value of the transmitted offset values in the control and positioning the automated storage and retrieval sys tem according to this average value is recommended In order to switch between Single and Double deep modes the control must report which rack depth is to be approached by means of a command If the position of the device and the defined target ...

Страница 26: ...device can also be controlled via the digital inputs and digital outputs Image recording is started via a trigger on the Sensor1 digital input The Sensor2 digital input is used to switch between the Single and Double deep modes The meas urement results are output via 4 digital outputs see 5 4 5 Control via digital inputs and digital outputs page 30 The device only delivers information on the direc...

Страница 27: ...ment Double deep 00 02 If a command is sent with the contents 00 02 then continuous measurement is switched on in Double deep mode Byte 1 Byte 2 Designation Command_H Command_L Function Command Example 3 00 02 Interpreta tion example 3 Switching on continuous measure ment Activating Double deep mode Command 4 Parameterization of diameter and distance 01 00 If the first byte Command_H equals 01 the...

Страница 28: ...OffY_H OffY_L TolX_H TolX_L TolY_H TolY_L Function Command Evaluate mode Relative offset to deviation X in 1 10 mm Relative offset to deviation Y in 1 10 mm Tolerance X in 1 10 mm Tolerance Y in 1 10 mm Example 5 02 00 09 00 64 FF 38 01 F4 01 2C Interpreta tion exam ple 5 Specialist single deep pa rameterization 9 target re gion best di ameter 100 10 mm 10 mm 200 10 mm 20 mm 500 10 mm 50 mm 300 10...

Страница 29: ...host 5 4 4 Answer from the device to the control As long as continuous measurement is switched on the device cyclically transmits measurement results to the control For each image 16 bytes are reported back to the control from the device Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Byte 6 Byte 7 Byte 8 Byte 9 Byte 10 Desig nation Status_H Status_L XposH_H XposH_L XposL_H XposL_L YposH_H YposH_L YposL_H Ypos...

Страница 30: ...and deactivates continuous measurement The sensor 2 digital input can be used to switch between the Single and Double deep modes Sensor 1 1 Continuous measurement is switched on 0 Continuous measurement is switched off Sensor 2 1 Double deep mode is activated 0 Single deep mode is activated The IN OUT 3 IN OUT 6 digital outputs give information about the current position of the positioning mark IN...

Страница 31: ...l and the target position of the automated storage and retrieval system is not transmitted The direction up down right left is also displayed on the device via LEDs 1 IN OUT 6 direction specification left 2 IN OUT 5 direction specification right 3 IN OUT 4 direction specification down 4 IN OUT 3 direction specification up The current position of the positioning mark is displayed in SOPASair in the...

Страница 32: ...efine hole or reflector which is to be used as the positioning mark If sev eral circles are located in the field of view of the device Selection of the correct hole or the correct reflector as the positioning mark Hiding of the non relevant im age section for several circles in the image in evaluation mode 9 to 13 Execute Auto gain Automatically set image brightness to make the positioning mark as...

Страница 33: ...lar this parameter ensures that they are still detected as positioning marks The detection of the positioning mark is optimized in relation to the deviat ing shape Double deep Switch between the Single and Dou ble deep modes Report to the device that settings for close range single deep or long range double deep are being made internally During set up this switch can be used to separately define i...

Страница 34: ...eter for example if the device is installed upside down in the system FTP server IP address Set IP address of the FTP server Define target address for image transmission PLC IP address Set IP address of control Define target address for data output to the control IP port InspectorP for TCP PLC is client Set IP ports for the device Define IP ports if the device is in the TCP server network Compact ...

Страница 35: ...s for position determina tion or to output positive feedback of successful positioning via the green feedback LED In Evaluation mode 9 to 13 the parameter defines the size of the target region whereby every thing outside the target region is ig nored during evaluation Inverting X Y axis Invert output values for the X axis and Y axis This parameter can be used for ex ample if the device is installe...

Страница 36: ...Phone 81 3 5309 2112 E Mail support sick jp Malaysia Phone 603 8080 7425 E Mail enquiry my sick com Mexico Phone 52 472 748 9451 E Mail mexico sick com Netherlands Phone 31 0 30 229 25 44 E Mail info sick nl New Zealand Phone 64 9 415 0459 0800 222 278 tollfree E Mail sales sick co nz Norway Phone 47 67 81 50 00 E Mail sick sick no Poland Phone 48 22 539 41 00 E Mail info sick pl Romania Phone 40 ...

Отзывы: