Installation
GMS815P-PS-2G/-3G · Supplementary Operating Instructions · 8016234 V2.0 · © SICK AG
15
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
3.2
Assembly
3.2.1
Ensuring ambient conditions
Quiet running
▸
Protect the device against heavy jolts and vibrations (limit values
Temperature
▸
Avoid enclosure exposure to direct sunlight.
▸
Maintain the allowable ambient temperature during operation (
).
Humidity
▸
Choose a dry installation location free from frost.
▸
Prevent moisture condensation – inside the device as well.
▸
Maintain the allowable relative air humidity (
).
Corrosive atmospheres
When the atmosphere at the installation location can contain corrosive gases:
▸
Install the Enclosure GMS815P-PS-2G/-3G in an outer housing (e.g. closed cabinet).
Purge the outer housing with a protective gas.
3.2.2
Fitting the enclosure
▸
Secure the enclosure on a structure that can safely carry the weight of the enclosure.
▸
Fit the enclosure so that the underside of the gas analyzer enclosure is more or less
horizontal (allowable offset
).
3.2.3
Removing transport safety devices
1
Open the upper section of the gas analyzer enclosure (
).
2
Take the foam inserts out of the enclosure.
▸
Recommendation:
Keep the foam inserts and use these every time the device is
transported.
3
Close the upper section of the gas analyzer enclosure again carefully; pay attention to
leak tightness (enclosure seal, screws on front door).
●
Dimensions
●
Weight